ArticleVascular health and risk management2023
Validation of the Withings BPM Core Device for Self-Blood Pressure Measurements in General Population According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard.
Article in Vascular health and risk management, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 4 citations in OpenAlex.
- Real-World Body Composition and Blood Pressure Changes Following Glucagon-Like Peptide 1 Receptor Agonist Initiation: A Causal Inference Study of Connected Device Data.Mayo Clinic proceedings. Digital health · 2026Article
- User experience of a combined 2-in-1 home ECG and blood pressure monitor: a qualitative study.The British journal of cardiology · 2026Article
- Accuracy of the Withings BPM Pro 2 (WPM07) Device for Self-Blood Pressure Measurements in the General Populations According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 2:2024) Protocol.Vascular health and risk management · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 3 institutions in 3 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Background: Most of the scientific societies recommend assessing the accuracy of electronic devices for blood pressure (BP) measurements using established validation protocol. Objective: To determine the accuracy of the BP measurements using the Withings BPM Core device in the general population according to the "Universal Standard (ISO 81060-2:2018/AMD 1:2020)". Methods: The Withings BPM Core is an oscillometric device measuring BP at the brachial level. The study was performed according to the "Universal Standard (ISO 81060-2:2018/AMD 1:2020) protocol" using the same-arm sequential BP measurement method. Subjects (n ≥ 85) fulfilling the age, gender, BP, and cuff distribution criteria of the protocol were included. Analysis was performed as required by the Universal protocol using Criterion 1 - differences between observers' mercury sphygmomanometer reference measurements and test device BP values (test versus reference) and their standard deviation (SD); and Criterion 2 - The SD of the mean BP differences between the test device and reference BP per subject. Results: Eighty-six subjects were selected, 85 of whom were included. The mean BP differences between the simultaneous two observers' measurements were -0.2 ± 2.1 mmHg for systolic BP (SBP) and 0.3 ± 2.1 mmHg for diastolic BP (DBP). For validation criterion 1, the mean difference ± SD between the reference and device BP values was -0.6 ± 4.8 mmHg for SBP and 0.1 ± 3.7 mmHg for DBP (≤5 ± 8 mmHg for both SBP and DBP). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 3.2/2.6 mmHg for SBP and DBP (≤6.91/6.95 mmHg). Conclusion: The results of this study showed that the Withings BPM Core oscillometric device for home BP measurement fulfilled the accuracy requirements of the (ISO 81060-2:2018/AMD 1:2020) Universal protocol in the general population.