ArticleIEEE transactions on bio-medical engineering2024
Measurement of Magnetically Induced Torque on Lightweight Medical Devices in the Magnetic Resonance Environment for ASTM F2213.
Article in IEEE transactions on bio-medical engineering, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed.
- Roadmap towards personalized approaches and safety considerations in non-ionizing radiation: from dosimetry to therapeutic and diagnostic applications.Physics in medicine and biology · 2026Review
- MRI Safety Considerations for Permanent Magnet Implants in Muscle.Journal of magnetic resonance imaging : JMRI · 2026Article
- Safety and Functional Integrity of Continuous Glucose Monitoring Sensors When Used During Radiologic Procedures Under High Exposure Conditions.Journal of diabetes science and technology · 2023Article
- Open Source MR-Safe Pneumatic Radial Inflow Motor and Encoder (PRIME): Design and Manufacturing Guidelines.... International Symposium on Medical Robotics. International Symposium on Medical Robotics · 2023Article
- ACPSEM position paper: the safety of magnetic resonance imaging linear accelerators.Physical and engineering sciences in medicine · 2023Article
- 7T MR Safety.Journal of magnetic resonance imaging : JMRI · 2021Article
- Simulated accuracy assessment of small footprint body-mounted probe alignment device for MRI-guided cryotherapy of abdominal lesions.Medical physics · 2020Article
- Magnetic resonance (MR) safety and compatibility of a novel iron bioresorbable scaffold.Bioactive materials · 2020Article
- Characterization and Control of a Pneumatic Motor for MR-conditional Robotic Applications.IEEE/ASME transactions on mechatronics : a joint publication of the IEEE Industrial Electronics Society and the ASME Dynamic Systems and Control Division · 2017Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
Abstract
objectiveUse of medical devices in the magnetic resonance environment is regulated by standards that include the ASTM-F2213 magnetically induced torque. This standard prescribes five tests. However, none can be directly applied to measure very low torques of slender lightweight devices such as needles.
methodsWe present a variant of an ASTM torsional spring method that makes a "spring" of 2 strings that suspend the needle by its ends. The magnetically induced torque on the needle causes it to rotate. The strings tilt and lift the needle. At equilibrium, the magnetically induced potential energy is balanced by the gravitational potential energy of the lift. Static equilibrium allows calculating the torque from the needle rotation angle, which is measured. Moreover, a maximum rotation angle corresponds to the maximum acceptable magnetically induced torque, under the most conservative ASTM acceptability criterion. A simple apparatus using the 2-string method is shown, it can be 3D printed, and the design files are shared.
resultsThe analytical methods were tested against a numeric dynamic model, showing perfect concordance. The method was then tested experimentally in 1.5T and 3T MRI with commercial biopsy needles. Numeric test errors were immeasurably small. Torques between 0.0001 Nm and 0.0018 Nm were measured in MRI with 7.7% maximum difference between tests. The cost to make the apparatus is 58USD and design files are shared.
conclusionThe apparatus is simple and inexpensive and provides good accuracy as well. SIGNIFICANCE: The 2-string method provides a solution to measure very low torques in the MRI.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials