ReviewClinical neuroradiology2023
Early Outcomes of the Pipeline Vantage Flow Diverter : A Multicentre Study.
Review in Clinical neuroradiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Initial Experience with the Pipeline Vantage Flow Diverter for Intracranial Aneurysms: A Systematic Review and Meta-Analysis.AJNR. American journal of neuroradiology · 2025Pooled it
- Midterm Results on Pipeline Vantage Embolization Device (MR-PVED): First Reported Experience in Asia.Asian journal of neurosurgery · 2026Article
- Safety and efficacy of the pipeline vantage flow diverter for the treatment of saccular aneurysms in the posterior cerebral circulation.The neuroradiology journal · 2026Article
- The DERIVO 2 Heal Embolization Device in the Treatment of Ruptured and Unruptured Intracranial Aneurysms: a Retrospective Multicenter Study.Clinical neuroradiology · 2025Article
- Multicenter Experience with the Pipeline Flex and Vantage with Shield Technology for Intracranial Aneurysm Treatment.AJNR. American journal of neuroradiology · 2024Observational
- Use of the Neuroform Atlas Stent or LVIS Jr Stent for Treatment of Unruptured Intracranial Aneurysms in Parent Arteries of <2 mm in Diameter: A Multicenter Experience.AJNR. American journal of neuroradiology · 2024Article
- Image Quality and Radiation Dose of Conventional and Wide-Field High-Resolution Cone-Beam Computed Tomography for Cerebral Angiography: A Phantom Study.Tomography (Ann Arbor, Mich.) · 2023Article
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe recently introduced Pipeline Vantage Embolization Device with Shield Technology is the fourth generation of Pipeline flow diverter devices. Due to the relatively high rate of intraprocedural technical complications, modifications were subsequently made to the device after a limited release of the device in 2020. This study aimed to evaluate the safety and efficacy of the modified version of this device.
methodsThis was a multicentre retrospective series. The primary efficacy endpoint was aneurysm occlusion in the absence of retreatment. The primary safety endpoint was any neurological morbidity or death. Ruptured and unruptured aneurysms were included in the study.
resultsA total of 52 procedures were performed for 60 target aneurysms. Treatment was performed on 5 patients with ruptured aneurysms. The technical success rate was 98%. The mean clinical follow-up time was 5.5 months. In patients presenting with unruptured aneurysms there were no deaths, 3 (6.4%) major complications and 7 (13%) minor complications. In the five patients presenting with subarachnoid haemorrhage there were 2 (40%) major complications with 1 (20%) of these resulting in death, and 1 (20%) minor complication. Of the patients 29 (56%) had undergone 6‑monthly postprocedural angiographic imaging with a mean time of 6.6 months demonstrating that 83% of patients had achieved adequate occlusion (RROC1/2) of the aneurysm.
conclusionsIn this non-industry-sponsored study, the occlusion rates and safety outcomes were similar to those seen in previously published studies with flow diverter devices and earlier generation Pipeline devices. Modifications to the device appear to have improved ease of deployment.
Indexed as
Identifiers
37378843What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.