A Comparison of Optical Biometers Used in Children for Myopia Control.
Article in Klinische Monatsblatter fur Augenheilkunde, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 12 citations in OpenAlex.
- Clinical Comparative Evaluation of Two New-Generation Optical Biometers: Intrasession Repeatability and Agreement of Biometric Parameters.Diagnostics (Basel, Switzerland) · 2026Article
- IMI-Instrumentation for Myopia Management.Investigative ophthalmology & visual science · 2025Review
- Inter-Examiner and Inter-Instrument Agreement of the Myopia Master with the IOL Master 700 and NVisionK 5001 in Myopic Children.Children (Basel, Switzerland) · 2025Article
- Effect of defocus incorporated multiple segments (DIMS) spectacle lenses on myopia progression in children: a retrospective analysis in a German real-life clinical setting.BMC ophthalmology · 2024Article
- Comparison of Eye Axial Length Measurements Taken Using Partial Coherence Interferometry and OCT Biometry.Vision (Basel, Switzerland) · 2024Article
- Age-matched analysis of axial length growth in myopic children wearing defocus incorporated multiple segments spectacle lenses.The British journal of ophthalmology · 2024Article
- The Repeatability and Agreement of Ocular Parameters Measured with the MYAH and Myopia Master Devices between Expert and Non-Expert Practitioners.Life (Basel, Switzerland) · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo assess the reproducibility (i.e., inter-device reliability) of the biometers Topcon MYAH, Oculus Myopia Master, and Haag-Streit Lenstar LS900 with the Carl Zeiss IOLMaster 700 and the intra-subject repeatability in myopic children in order to reliably detect axial growth for myopia management.
methodsTwenty-two myopic children (11.1 ± 2.4 yr) with a spherical equivalent of - 3.53 ± 2.35 D were examined with each of the biometers to assess axial length (AL) and corneal parameters (steepK, flatK, meanK, vectors J0, J45), and16 of these children agreed to a second round of measurements. Reproducibility of the first measurements between the IOLMaster and every other biometer was assessed employing a Bland-Altman approach and paired Student's t-test. Repeatability was assessed as intra-subject standard deviation and was used to estimate the minimum time interval required between two AL measurements to reliably detect axial growth of an eye of at least 0.1 mm/year.
resultsRepeatability for AL measurements was as follows: IOLMaster: 0.05 mm, Myopia Master: 0.06 mm, Myah: 0.06 mm, Lenstar: 0.04 mm; the respective minimal time interval for axial growth assessment in myopia management was estimated as 5.6, 6.6, 6.7, and 5.0 months, respectively. Best reproducibility of the AL measurement was found between IOLMaster and Lenstar [95% Limits of Agreement (LoA) for reproducibility - 0.06 to 0.02]. As regards to the measured means, Lenstar gave measures of AL that were longer than with the IOLMaster by 0.02 mm (p < 0.001). Myopia Master measures of meanK were significantly lower (by 0.21 D with p < 0.001) than the values from the IOLMaster. As for J0, all biometers deviated significantly from IOLMaster measurements (p < 0.05).
conclusionGenerally good agreement was observed between all the biometers. When assessing myopia progression in children, a time frame of at least 6 months between the AL measurements is advisable in order to reliably determine any deviation from a normal growth pattern.