ArticleJournal of neurointerventional surgery2024
High variability in physician estimations of flow-diverting stent deployment versus PreSize Neurovascular software simulation: a comparison study.
Article in Journal of neurointerventional surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 6 citations in OpenAlex.
- Stroke-Associated Infections after Mechanical Thrombectomy: Analysis of the Stroke-Associated INfection afTer Mechanical Thrombectomy (SAINT-MT) Study.Cerebrovascular diseases extra · 2026Article
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors at 7 institutions in 4 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPhysician variablity in preoperative planning of endovascular implant deployment and associated inaccuracies have not been documented. This study aimed to quantify the variability in accuracy of physician flow diverter (FD) planning and directly compares it with PreSize Neurovascular (Oxford Heartbeat Ltd) software simulations.
methodsEight experienced neurointerventionalists (NIs), blinded to procedural details, were provided with preoperative 3D rotational angiography (3D-RA) volumetric data along with images annotated with the distal landing location of a deployed Surpass Evolve (Stryker Neurovascular) FD from 51 patient cases. NIs were asked to perform a planning routine reflecting their normal practice and estimate the stent's proximal landing using volumetric data and the labeled dimensions of the FD used. Equivalent deployed length estimation was performed using PreSize software. NI- and software-estimated lengths were compared with postprocedural observed deployed stent length (control) using Bland-Altman plots. NI assessment agreement was assessed with the intraclass correlation coefficient (ICC).
resultsThe mean accuracy of NI-estimated deployed FD length was 81% (±15%) versus PreSize's accuracy of 95% (±4%), demonstrating significantly higher accuracy for the software (p<0.001). The mean absolute error between estimated and control lengths was 4 mm (±3.5 mm, range 0.03-30.2 mm) for NIs and 1 mm (±0.9 mm, range 0.01-3.9 mm) for PreSize. No discernable trends in accuracy among NIs or across vasculature and aneurysm morphology (size, vessel diameter, tortuousity) were found.
conclusionsThe study quantified experienced physicians' significant variablity in predicting an FD deployment with current planning approaches. In comparison, PreSize-simulated FD deployment was consistently more accurate and reliable, demonstrating its potential to improve standard of practice.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.