Observational studyKnee surgery, sports traumatology, arthroscopy : official journal of the ESSKA2023
Intraoperative control by Schanz-screws is inaccurate to achieve the exact amount of correction in de-rotational osteotomies.
Observational study in Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe accuracy of intraoperative control of correction commonly is achieved by K-wires or Schanz-screws in combination with goniometer in de-rotational osteotomies. The purpose of this study is to investigate the accuracy of intraoperative torsional control in de-rotational femoral and tibial osteotomies. It is hypothesized, that intraoperative control by Schanz-screws and goniometer in de-rotational osteotomies around the knee is a safe and well predictable method to control the surgical torsional correction intraoperatively.
methods55 consecutive osteotomies around the knee joint were registered, 28 femoral and 27 tibial. The indication for osteotomy was femoral or tibial torsional deformity with the clinical occurrence of patellofemoral maltracking or PFI. Pre- and postoperative torsions were measured according to the method of Waidelich on computed tomography (CT) scan. The scheduled value of torsional correction was defined by the surgeon preoperatively. Intraoperative control of torsional correction was achieved by 5 mm-Schanz-screws and goniometer. The measured values of torsional CT scan were compared to the preoperative defined and intended values and deviation was calculated separately for femoral and tibial osteotomies.
resultsThe surgeon's intraoperative measured mean value of correction in all osteotomies was 15.2° (SD 4.6; range 10-27), whereas the postoperatively measured mean value on CT scan was 15.6 (6.8; 5.0-28.5). Intraoperatively the femoral mean value measured 17.9° (4.9; 10-27) and 12.4° (1.9; 10-15) for the tibia. Postoperatively the mean value for femoral correction was 19.8 (5.5; 9.0-28.5) and 11.3 (5.0; 5.0-26.0) for tibial correction. When considering a deviation of plus or minus 3° to be acceptable femorally 15 osteotomies (53.6%) and tibially 14 osteotomies (51.9%) fell within these limits. Nine femoral cases (32.1.%) were overcorrected, four cases undercorrected (14.3%). Four tibial cases of overcorrection (14.8%) and 9 tibial cases of undercorrection (33.3%) were observed. However, the observed difference between femur and tibia regarding the distribution of cases between the three groups did not reach significance. Moreover, there was no correlation between the extent of correction and the deviation from the intended result.
conclusionThe use of Schanz-screws and goniometer in de-rotational osteotomies as an intraoperative control of correction is an inaccurate method. Every surgeon performing derotational osteotomies must consider this and include postoperative torsional measurement in his postoperative algorithm until new tools or devices are available to guarantee a better intraoperative accuracy of torsional correction. STUDY
designObservational study. LEVEL OF EVIDENCE: III.
Indexed as
Identifiers
37329368What OpenQuestion holds
Registered trials