ArticleJournal of applied clinical medical physics2023
Comprehensive MR imaging QA of 0.35 T MR-Linac using a multi-purpose large FOV phantom: A single-institution experience.
Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed, 6 citations in OpenAlex.
- Acceptance testing of a 0.35 T MR-Linac: procedures, QA baseline, and system limitations.Journal of applied clinical medical physics · 2026Article
- Evaluation of QUASAR Insight Phantom for daily imaging QA of MRgRT linacs.Journal of applied clinical medical physics · 2026Article
- Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner.Physics and imaging in radiation oncology · 2024Article
- Receive coil quality assurance procedure and automated analysis for ViewRay MRIdian MR-Linac.Journal of applied clinical medical physics · 2024Article
- Comprehensive MR imaging QA of 0.35 T MR-Linac using a multi-purpose large FOV phantom: A single-institution experience.Journal of applied clinical medical physics · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 2 institutions in 2 countries.
Funding
Abstract
purposeMagnetic resonance-guided radiotherapy (MRgRT) is desired for the treatment of diseases in the abdominothoracic region, which has a broad imaging area and continuous motion. To ensure accurate treatment delivery, an effective image quality assurance (QA) program, with a phantom that covers the field of view (FOV) similar to a human torso, is required. However, routine image QA for a large FOV is not readily available at many MRgRT centers. In this work, we present the clinical experience of the large FOV MRgRT Insight phantom for periodic daily and monthly comprehensive magnetic resonance imaging (MRI)-QA and its feasibility compared to the existing institutional routine MRI-QA procedures in 0.35 T MRgRT.
methodsThree phantoms; ViewRay cylindrical water phantom, Fluke 76-907 uniformity and linearity phantom, and Modus QA large FOV MRgRT Insight phantom, were imaged on the 0.35 T MR-Linac. The measurements were made in MRI mode with the true fast imaging with steady-state free precession (TRUFI) sequence. The ViewRay cylindrical water phantom was imaged in a single-position setup whereas the Fluke phantom and Insight phantom were imaged in three different orientations: axial, sagittal, and coronal. Additionally, the phased array coil QA was performed using the horizontal base plate of the Insight phantom by placing the desired coil around the base section which was compared to an in-house built Polyurethane foam phantom for reference.
resultThe Insight phantom captured image artifacts across the entire planar field of view, up to 400 mm, in a single image acquisition, which is beyond the FOV of the conventional phantoms. The geometric distortion test showed a similar distortion of 0.45 ± 0.01 and 0.41 ± 0.01 mm near the isocenter, that is, within 300 mm lengths for Fluke and Insight phantoms, respectively, but showed higher geometric distortion of 0.8 ± 0.4 mm in the peripheral region between 300 and 400 mm of the imaging slice for the Insight phantom. The Insight phantom with multiple image quality features and its accompanying software utilized the modulation transform function (MTF) to evaluate the image spatial resolution. The average MTF values were 0.35 ± 0.01, 0.35 ± 0.01, and 0.34 ± 0.03 for axial, coronal, and sagittal images, respectively. The plane alignment and spatial accuracy of the ViewRay water phantom were measured manually. The phased array coil test for both the Insight phantom and the Polyurethane foam phantoms ensured the proper functionality of each coil element.
conclusionThe multifunctional large FOV Insight phantom helps in tracking MR imaging quality of the system to a larger extent compared to the routine daily and monthly QA phantoms currently used in our institute. Also, the Insight phantom is found to be more feasible for routine QA with easy setup.