Formaldehyde Reduction in an Operating Room Setting: Comparison of a Catalytic Surgical Vacuum Device With a Traditional Smoke Evacuator.

Gregory T Carroll, David L Kirschman

Open access · diamondAbstract read
In one paragraph

Article in Cureus, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact, top 77% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 0 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors at 1 institution in 1 country.

Gregory T CarrollScientific Affairs, Aerobiotix Inc, Miamisburg, USA.
David L KirschmanScientific Affairs, Aerobiotix Inc, Miamisburg, USA.
Aerobotix (United States) · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Introduction Electrosurgery exposes healthcare workers to volatile organic compounds (VOCs) including formaldehyde. Adopting electrosurgical devices that catalytically transform formaldehyde to benign substances has the potential to improve safety in surgical settings. Materials and methods We compared the efficiency of formaldehyde removal of two medical devices. The first was a novel surgical vacuum (SV) device containing ultra-low particulate air (ULPA) filtration, activated carbon and catalytic transition metal oxide. The second was a commonly utilized handpiece evacuator (HE) that contained only mechanical filtration and activated carbon granules. Both devices were exposed to formalin vapor. Results The time weighted average (TWA), median and peak concentrations of detected formaldehyde at the outflow of the SV unit were 90% lower than the corresponding values detected at the outflow of the HE device (p = 0.0034). When catalytic material was added to the HE device, the detected formaldehyde concentration at the outflow was reduced by 55% (p = 2.9 x 10

Indexed as

activated carbonair qualitycatalysiselectrosurgeryformaldehydemedical devicesoperating roomportable filtrationsurgical smokevocs

Identifiers

PMID37303407
PMCPMC10253242
OpenAlexW4376113685