Evidence map›Paper›PMID 37287131›Full record

ArticleJournal of applied clinical medical physics2023

Comparing brass mesh to tissue equivalent bolus materials for volumetric modulated arc therapy chest wall irradiation.

Timothy D Keiper, Kelly Kisling, Patricia Hua, Ryan P Manger

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Timothy D KeiperDepartment of Radiation Medicine and Applied Sciences, Moores Cancer Center, University of California San Diego, La Jolla, California, USA.
Kelly KislingDepartment of Radiation Medicine and Applied Sciences, Moores Cancer Center, University of California San Diego, La Jolla, California, USA.
Patricia HuaDepartment of Radiation Medicine and Applied Sciences, Moores Cancer Center, University of California San Diego, La Jolla, California, USA.
Ryan P MangerDepartment of Radiation Medicine and Applied Sciences, Moores Cancer Center, University of California San Diego, La Jolla, California, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo compare the superficial dose when using brass mesh bolus (BMB), no bolus, or 3 mm tissue-equivalent bolus with a pseudo-flash volumetric modulated arc therapy (VMAT) breast treatment planning technique.

methodsTwo different beam arrangements for right-sided irradiation and one beam arrangement for bilateral irradiation were planned on an inhomogeneous thorax phantom in accordance with our clinical practice for VMAT postmastectomy radiotherapy (PMRT). Plans were optimized using pseudo-flash and representative critical organ optimization structures were used to shape the dose. Plans were delivered without bolus, with 3 mm tissue-equivalent bolus (TEB), or with one-layer BMB. Optically stimulated luminescence dosimeter (OSLD) and radiochromic film measurements were taken and analyzed to determine the superficial dose in each case and the relative enhancement from the no bolus delivery.

resultsSuperficial dose measured with OSLDs was found to be 76.4 ± 4.5%, 103.0 ± 6.1%, and 98.1 ± 5.8% of prescription for no physical bolus (NB), TEB, and BMB, respectively. Superficial dose was observed to increase from lateral to medial points when measured with film. However, the relative increase in superficial dose from NB was consistent across the profile with an increase of 43 ± 2.1% and 34 ± 3.3% of prescription for TEB and BMB, respectively. The results are in good agreement with expectations from the literature and the experience with tangential radiotherapy.

conclusionThree millimeter TEB and one-layer BMB were shown to provide similar enhancement to the superficial dose compared to delivery without bolus. BMB, which does not significantly affect dose at depth and is more conformal to the patient surface, is an acceptable alternative to 3 mm TEB for chest wall PMRT patients treated with pseudo-flash PMRT.

Indexed as

Breast NeoplasmsRadiotherapy, Intensity-ModulatedThoracic WallCopperFemaleHumansMastectomyRadiotherapy DosageRadiotherapy Planning, Computer-AssistedZincbrassCopperZincbrass bolusbrass meshVMAT breast

Identifiers

PMID37287131
PMCPMC10476985

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.