Evidence map›Paper›PMID 37278646›Full record

ArticleJournal of applied clinical medical physics2023

Commissioning, clinical implementation, and initial experience with a new brain tumor treatment package on a low-field MR-linac.

Karen Chin Snyder, Weihua Mao, Joshua P Kim, Justine Cunningham, Indrin J Chetty, Salim M Siddiqui, Parag Parikh, Jennifer Dolan

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.6field-weighted citation impact, top 17% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 7 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 1 institution in 1 country.

Karen Chin SnyderDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Weihua MaoDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Joshua P KimDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Justine CunninghamDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Indrin J ChettyDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Salim M SiddiquiDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Parag ParikhDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Jennifer DolanDepartment of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.
Henry Ford Health System · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

To evaluate the image quality, dosimetric properties, setup reproducibility, and planar cine motion detection of a high-resolution brain coil and integrated stereotactic brain immobilization system that constitute a new brain treatment package (BTP) on a low-field magnetic resonance imaging (MRI) linear accelerator (MR-linac). Image quality of the high-resolution brain coil was evaluated with the 17 cm diameter spherical phantom and the American College of Radiology (ACR) Large MRI Phantom. Patient imaging studies approved by the institutional review board (IRB) assisted in selecting image acquisition parameters. Radiographic and dosimetric evaluation of the high-resolution brain coil and the associated immobilization devices was performed using dose calculations and ion chamber measurements. End-to-end testing was performed simulating a cranial lesion in a phantom. Inter-fraction setup variability and motion detection tests were evaluated on four healthy volunteers. Inter-fraction variability was assessed based on three repeat setups for each volunteer. Motion detection was evaluated using three-plane (axial, coronal, and sagittal) MR-cine imaging sessions, where volunteers were asked to perform a set of specific motions. The images were post-processed and evaluated using an in-house program. Contrast resolution of the high-resolution brain coil is superior to the head/neck and torso coils. The BTP receiver coils have an average HU value of 525 HU. The most significant radiation attenuation (3.14%) of the BTP, occurs through the lateral portion of the overlay board where the high-precision lateral-profile mask clips attach to the overlay. The greatest inter-fraction setup variability occurred in the pitch (average 1.08 degree) and translationally in the superior/inferior direction (average 4.88 mm). Three plane cine imaging with the BTP was able to detect large and small motions. Small voluntary motions, sub-millimeter in magnitude (maximum 0.9 mm), from motion of external limbs were detected. Imaging tests, inter-fraction setup variability, attenuation, and end-to-end measurements were quantified and performed for the BTP. Results demonstrate better contrast resolution and low contrast detectability that allows for better visualization of soft tissue anatomical changes relative to head/neck and torso coil systems.

Indexed as

Brain NeoplasmsBrainHumansImaging, Three-DimensionalMagnetic Resonance ImagingPhantoms, ImagingReproducibility of Resultsbraininter-fraction motionMR coilMRgRT