Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer.

Alberto Domínguez-Vicent, Abinaya Priya Venkataraman, Amanda Dalin, Rune Brautaset, Robert Montés-Micó

Open access · diamondAbstract read
In one paragraph

Article in Eye and vision (London, England), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
2.3field-weighted citation impact, top 12% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed, 10 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Review
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 2 countries.

Alberto Domínguez-VicentDivision of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, 171 77, Stockholm, Sweden. alberto.dominguez.vicent@ki.se.
Abinaya Priya VenkataramanDivision of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, 171 77, Stockholm, Sweden.
Amanda DalinDivision of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, 171 77, Stockholm, Sweden.
Rune BrautasetDivision of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, 171 77, Stockholm, Sweden.
Robert Montés-MicóOptics and Optometry and Vision Sciences Department, University of Valencia, Valencia, Spain.
Karolinska Institutet · SEUniversitat de València · ES

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundTo evaluate the repeatability of a fully automated swept-source optical coherence tomography (SS-OCT) and its agreement with an optical low coherence reflectometry (OLCR) for several biometric parameters.

methodsIn this study, 74 eyes of 74 patients were measured using the Eyestar 900 SS-OCT and Lenstar LS 900 OLCR. Flat keratometry (K1) and steep keratometry (K2), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and axial length (AL) were measured three times with each device. The repeatability was analyzed with the intrasubject standard deviation, coefficient of variability (CoV), and coefficient of repeatability (CoR) for each instrument. The agreement between the instruments was evaluated with Bland-Altman analysis.

resultsK1, K2 and CCT CoV values were < 0.2%, < 0.4% and < 0.55%, respectively. Higher CoV values were found for ACD and LT ranging from 0.56% to 1.74%. The lowest CoV values were found for the AL measurements (0.03% and 0.06% for the Eyestar 900 and the Lenstar LS 900, respectively). AL measurements provided the highest repeatability, measured with both CoV and CoR values, and the CCT was the parameter with the lowest repeatability. The CCT and LT measurements were statistically significant between the two biometers (P < 0.001). The interval of the limits of agreement was < 0.6 D for K1 and K2, 15.78 µm for CCT, 0.21 mm for ACD, 0.34 mm for LT, and 0.08 mm for AL.

conclusionsBoth biometers provide repeatable measurements for the different parameters analyzed and can be used interchangeably.

Indexed as

AgreementLow-coherence interferometryOptical biometryRepeatabilitySwept-source optical coherence tomography

Identifiers

PMID37264436
PMCPMC10236819
OpenAlexW4379055022

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.