ArticleInternational journal of computer assisted radiology and surgery2023
Feasibility of tracked ultrasound registration for pelvic-abdominal tumor navigation: a patient study.
Article in International journal of computer assisted radiology and surgery, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT05637359 (Tracked Ultrasound for Patient Registration in Surgical Navigation During Abdominal Cancer Surgery), which is not on this map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Tracked Ultrasound for Patient Registration in Surgical Navigation During Abdominal Cancer Surgery
Who cites it
6 citing papers in PubMed, 8 citations in OpenAlex.
- Real-time intraoperative ultrasound registration for accurate surgical navigation in patients with pelvic malignancies.International journal of computer assisted radiology and surgery · 2025Trial
- Electromagnetic Surgical Navigation for Accurate Pediatric Tumor Localization: A Preclinical Animal Cadaver Study.Cancers · 2026Article
- Real-time surgical navigation for intraoperative lymph node localization during targeted robot-assisted pelvic and retroperitoneal lymph node dissection.Surgical endoscopy · 2026Article
- Performance of a tracked ultrasound navigation setup for pediatric tumor surgery: a phantom study.International journal of computer assisted radiology and surgery · 2026Article
- A Miniature A-Mode Ultrasound System for Noninvasive Bone Surface Point Cloud Acquisition.Annals of biomedical engineering · 2026Article
- Surgical navigation for targeted retroperitoneal lymph-node removal: a randomised, controlled, phase 3 trial.EClinicalMedicine · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 1 country.
Funding
Abstract
purposeSurgical navigation techniques can guide surgeons in localizing pelvic-abdominal malignancies. For abdominal navigation, accurate patient registration is crucial and is generally performed using an intra-operative cone-beam CT (CBCT). However, this method causes 15-min surgical preparation workflow interruption and radiation exposure, and more importantly, it cannot be repeated during surgery to compensate for large patient movement. As an alternative, the accuracy and feasibility of tracked ultrasound (US) registration are assessed in this patient study.
methodsPatients scheduled for surgical navigation during laparotomy of pelvic-abdominal malignancies were prospectively included. In the operating room, two percutaneous tracked US scans of the pelvic bone were acquired: one in supine and one in Trendelenburg patient position. Postoperatively, the bone surface was semiautomatically segmented from US images and registered to the bone surface on the preoperative CT scan. The US registration accuracy was computed using the CBCT registration as a reference and acquisition times were compared. Additionally, both US measurements were compared to quantify the registration error caused by patient movement into Trendelenburg.
resultsIn total, 18 patients were included and analyzed. US registration resulted in a mean surface registration error of 1.2 ± 0.2 mm and a mean target registration error of 3.3 ± 1.4 mm. US acquisitions were 4 × faster than the CBCT scans (two-sample t-test P < 0.05) and could even be performed during standard patient preparation before skin incision. Patient repositioning in Trendelenburg caused a mean target registration error of 7.7 ± 3.3 mm, mainly in cranial direction.
conclusionUS registration based on the pelvic bone is accurate, fast and feasible for surgical navigation. Further optimization of the bone segmentation algorithm will allow for real-time registration in the clinical workflow. In the end, this would allow intra-operative US registration to correct for large patient movement.
trial registrationThis study is registered in ClinicalTrials.gov (NCT05637359).
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.