Evidence map›Paper›PMID 37224472›Full record

ArticleRetinal cases & brief reports2023

Quantifying Head Position Following Vitrectomy Using a 3-Dimensional Position Sensor Embedded in a Novel Ophthalmic Shield: A Pilot Study.

Raziyeh Mahmoudzadeh, Abtin Shahlaee, Rebecca R Soares, Samir N Patel, David Xu, Omesh P Gupta, Allen Chiang

Abstract read
In one paragraph

Article in Retinal cases & brief reports, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Raziyeh MahmoudzadehWills Eye Hospital, Mid Atlantic Retina, Thomas Jefferson University, Philadelphia, PA.
Abtin Shahlaee
Rebecca R Soares
Samir N Patel
David Xu
Omesh P Gupta
Allen Chiang

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo assess head position following pars plana vitrectomy (PPV) using a novel low-energy, non-intrusive 3-D position sensor platform.

methodsIn this prospective non-randomized interventional case series, a low-energy Bluetooth smart sensor housed within a novel eye shield recorded 3-D positional data every five minutes. The device was placed on the patient immediately after PPV and data was retrieved at the postoperative day 1 visit. Readings were processed by vector analysis into 4 groups, stratified by the angle of deviation away from a completely prone head position. The primary outcome was the angle between vectors.

resultsTen patients were enrolled in this pilot study. The mean (SD) age was 57.5 (17.4). A total of 2318 readings with a mean (SD) of 231.8 (26.8) readings per patient were obtained. The mean (SD) number of readings were 132.9 (34.7) while awake and 98.9 (27.9) during sleep. Of total readings, only 11.7% fell into group 1 while a majority of readings fell in group 2 (52.4%), and group 3 (32.4%), and only 3.5% fell into group 4. Positional deviation increased during sleep hours, with about 46.8% of readings at a reclined angle (group 3) and 4.9% supine, which was 21.6% and 2.5% respectively during the wake time (p<0.001, p=0.002, respectively).

conclusionIn this pilot study, a non-intrusive wireless 3-D position sensor-shield platform was tolerated well and capable of capturing positional data. Adherence to face-down positioning was low and positional deviation increased significantly while sleeping.

Identifiers

PMID37224472
PMCPMC11027976

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.