Nano-Indentation to Determine Mechanical Properties of Intraocular Lenses: Evaluating Penetration Depth, Material Stiffness, and Elastic Moduli.
Article in Ophthalmology and therapy, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 3 citations in OpenAlex.
- Scanning electron microscopy of artificial corneal endothelial keratoplasty implant.Journal of cataract and refractive surgery · 2026Article
- Miri 4-points marker: a low-cost marking device for flanged intrascleral intraocular lens fixation with modified Yamane technique.BMC ophthalmology · 2026Article
- Extracellular Matrix Viscoelasticity: A Dynamic Regulator of Cellular Behavior.Annals of biomedical engineering · 2025Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionIntraocular lenses (IOL) should remain in the eye for life after implantation into the capsular bag during cataract surgery. The material must meet various requirements. It is crucial that the material has the best biocompatibility, and it should be flexible and soft for best possible implantation process but also sufficiently stable and stiff for good centering in the eye and posterior capsule opacification prevention.
methodsIn this laboratory experiment, we used nano-indentation for the mechanical assessment of three hydrophobic acrylic (A, B, C), three hydrophilic acrylic (D, E, F), and one silicone (G) intraocular lens. We wanted to determine whether some react more sensitively to touching/handling than others. The indentation elastic modulus and the creep were obtained from the force displacement curve. For measuring penetration depth and testing of possible damage to the intraocular lenses, the samples were measured at room temperature. A 200-µm-diameter ruby spherical tipped indenter was used for all the tests. Indentations were made to three different maximum loads, namely 5 mN (milli Newton), 15 mN, and 30 mN and repeated three times.
resultsThe lowest penetration depth (12 µm) was observed with IOL B. However, IOL A, D, and F showed similar low penetration depths (20, 18, and 23 µm, respectively). Lenses C and E showed slightly higher penetration depths of 36 and 39 µm, respectively. The silicone lens (G) showed the greatest penetration depth of 54.6 µm at a maximum load of 5 mN. With higher maximal loads (15 and 30 mN) the penetration depth increased significantly. Lens C, however, showed the same results at both 15 and 30 mN with no increase of penetration depth. This seems to fit well with the material and manufacturing process of the lens (lathe-cut). During the holding time of 30 s at constant force all six acrylic lenses showed a significant increase of the creep (C
conclusionIt was found that results correlate very well with the water content of the material in the first place. The manufacturing process (molded versus lathe-cut) seems to play another important role. Since all included acrylic lenses are very similar, it was not surprising that the measured differences are marginal. Even though hydrophobic materials with lower water content showed higher relative stiffness, penetration and defects can also occur with these. The surgeon and scrub nurse should always be aware that macroscopic changes are difficult to detect but that defects could theoretically lead to clinical effects. The principle of not touching the center of the IOL optic at any time should be taken seriously.