ArticleEuropean radiology2023
Safety and performance of MR-conditional pacing systems with automated MRI mode at 1.5 and 3 Tesla.
Article in European radiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
12 authors.
Funding
Abstract
objectivesTo evaluate at 1.5 and 3 T MRI the safety and performance of trademarked ENO
methodsA total of 267 implanted patients underwent MRI examination (brain, cardiac, shoulder, cervical spine) at 1.5 (n = 126) or 3 T (n = 141). MRI-related device complications, lead electrical performances stability at 1-month post-MRI, proper functioning of the automated MRI mode and image quality were evaluated.
resultsFreedom from MRI-related complications at 1 month post-MRI was 100% in both 1.5 and 3 T arms (both p < 0.0001). The stability of pacing capture threshold was respectively at 1.5 and 3 T (atrial:: 98.9% (p = 0.001) and 100% (p < 0.0001); ventricular: both 100% (p < 0001)). The stability of sensing was respectively at 1.5 and 3 T (atrial: 100% (p = 0.0001) and 96.9% (p = 0.01); ventricular: 100% (p < 0.0001) and 99.1% (p = 0.0001)). All devices switched automatically to the programmed asynchronous mode in the MRI environment and to initially programmed mode after the MRI exam. While all MR examinations were assessed as interpretable, artifacts deteriorated a subset of examinations including mostly cardiac and shoulder ones.
conclusionThis study demonstrates the safety and electrical stability of ENO CLINICAL RELEVANCE STATEMENT: ENO KEY POINTS: • Patients implanted with an MRI conditional cardiac pacemaker can be safely scanned under 1.5 or 3 Tesla MRI with preserved interpretability. • Electrical parameters of the MRI conditional pacing system remain stable after a 1.5 or 3 Tesla MRI scan. • The automated MRI mode enabled the automatic switch to asynchronous mode in the MRI environment and to initial settings after the MRI scan in all patients.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.