Medical Imaging Compatibility of Magnesium- and Iron-Based Bioresorbable Flow Diverters.
Article in AJNR. American journal of neuroradiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Bioresorbable flow-diverters for cerebral aneurysms: a systematic review.Neuroradiology · 2026Pooled it
- Degradation of biodegradable orthopedic screws does (not) elevate tissue heating under Magnetic Resonance Imaging: Implications for next generation implants.Bioactive materials · 2026Article
- Development and translation of biodegradable metal stents: from heart to brain.Regenerative biomaterials · 2026Review
- Magnesium-based bioresorbable flow diverter for intracranial aneurysms: a pilot study of biocompatibility and bioresorption in a rabbit vascular model.Journal of neurointerventional surgery · 2025Article
- Evaluation of FeMnN alloy bioresorbable flow diverting stents in the rabbit abdominal aorta.Bioactive materials · 2025Article
- Evaluation of FeMnN alloy bioresorbable flow diverters in the rabbit elastase induced aneurysm model.Frontiers in bioengineering and biotechnology · 2025Article
- Radiopaque FeMnN-Mo composite drawn filled tubing wires for braided absorbable neurovascular devices.Bioactive materials · 2024Article
- Absorbable metal stents for vascular use in pediatric cardiology: progress and outlook.Frontiers in cardiovascular medicine · 2024Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
12 authors.
Funding
Abstract
background and purposeBioresorbable flow diverters are under development to mitigate complications associated with conventional flow-diverter technology. One proposed advantage is the ability to reduce metal-induced artifacts in follow-up medical imaging. In the current work, the medical imaging compatibility of magnesium- and iron-based bioresorbable flow diverters is assessed relative to an FDA-approved control in phantom models. MATERIALS AND
methodsBioresorbable flow diverters, primarily composed of braided magnesium or antiferromagnetic iron alloy wires, were compared with an FDA-approved control flow diverter. The devices were assessed for MR imaging safety in terms of magnetically induced force and radiofrequency heating using 1.5T, 3T, and 7T field strength clinical scanners. The devices were deployed in phantom models, and metal-induced image artifacts were assessed in the 3 MR imaging scanners and a clinical CT scanner following clinical scan protocols; device visibility was assessed under fluoroscopy.
resultsThe magnesium-based bioresorbable flow diverter, iron-based bioresorbable flow diverter, and the control device all demonstrated MR imaging safety in terms of magnetically induced force and radiofrequency heating at all 3 field strengths. The bioresorbable flow diverters did not elicit excessive MR imaging artifacts at any field strength relative to the control. Furthermore, the bioresorbable flow diverters appeared to reduce blooming artifacts in CT relative to the control. The iron-based bioresorbable flow diverter and control device were visible under standard fluoroscopy.
conclusionsWe have demonstrated the baseline medical imaging compatibility of magnesium and antiferromagnetic iron alloy bioresorbable flow diverters. Future work will evaluate the medical imaging characteristics of the bioresorbable flow diverters in large-animal models.