Evidence map›Paper›PMID 37143316›Full record

ArticleJournal of applied clinical medical physics2023

Application of repeat image analysis to radiation therapy imaging modalities as a quality improvement tool for image guided radiotherapy.

Daniella Chee, Lesley Buckley

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.7field-weighted citation impact, top 33% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 2 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors at 2 institutions in 2 countries.

Daniella CheeRadiation Therapy Department, Princess Margaret Hospital, Toronto, Ontario, Canada.
Lesley BuckleyMedical Physics Department, The Ottawa Hospital, Ottawa, Ontario, Canada.
Princess Margaret Hospital · NZUniversity of Ottawa · CA

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundRepeat images contribute to excess patient dose and workflow inefficiencies and can be analyzed to identify potential areas for improvement within a program. Although routinely used in diagnostic imaging, repeat image analysis is not widely used in radiation therapy imaging, despite the role of imaging in the delivery of precise radiation treatments. PURPOSE: Repeat image analysis was performed for on-board cone beam CT imagers and CT simulators within a radiation therapy department. Both the rate of repeat images and the reasons for the repeat images were analyzed.

methodsData from nine conventional linear accelerators and three CT simulators were analyzed retrospectively over a 5-month period. Repeated images that were not part of the standard of care were considered repeat images. The repeat rate was expressed as the number of repeat scans as percentage of the total number of scans performed. The reasons for the repeats were collected and classified as either patient preparation, patient setup, patient motion, or machine error. These reasons were further classified into sub-categories.

resultsThe overall repeat rate across the linear accelerators was 3.3%, with a maximum of 5% on any single machine. The repeat rate for the three CT simulators was 1.5%. The most common reasons for repeat images were patient preparation (incorrect bladder or rectal filling) and patient setup or positioning. Greater positioning challenges led to higher repeat rates on units that treat a large number of breast patients, palliative patients, or pediatric patients.

conclusionsRepeat image analysis can be applied within a radiation therapy department. Establishing baseline repeat rates and analyzing reasons for the repeat images can identify opportunities for improvements in terms of patient dose reduction and workflow efficiency for the program. Periodic repeat image analysis also permits monitoring the program for changes and for comparison against rates at other institutions.

Indexed as

Radiotherapy, Image-GuidedChildCone-Beam Computed TomographyHumansQuality ImprovementRadiotherapy Planning, Computer-AssistedRetrospective StudiesIGRTradiation therapyrepeat image analysis