ArticleJournal of applied clinical medical physics2023
Sensitivity and specificity of Varian Halcyon's portal dosimetry for plan-specific pre-treatment QA.
Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 5 citations in OpenAlex.
- A component-wise commissioning and validation framework for non-transit electronic portal imaging device patient-specific quality assurance.Physics and imaging in radiation oncology · 2026Article
- Assessment of the backscattered skin dose from the beam stopper on Halcyon and Ethos radiotherapy systems.Journal of applied clinical medical physics · 2025Article
- Sensitivity and specificity of Varian Halcyon's portal dosimetry for plan-specific pre-treatment QA.Journal of applied clinical medical physics · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeDeveloped as a plan-specific pre-treatment QA tool, Varian portal dosimetry promises a fast, high-resolution, and integrated QA solution. In this study, the agreement between predicted fluence and measured cumulative portal dose was determined for the first 140 patient plans at our Halcyon linear accelerator. Furthermore, the capability of portal dosimetry to detect incorrect plan delivery was compared to that of a common QA phantom. Finally, tolerance criteria for verification of VMAT plan delivery with Varian portal dosimetry were derived.
methodsAll patient plans and the corresponding verification plans were generated within the Eclipse treatment planning system. Four representative plans of different treatment sites (prostate, prostate with lymphatic drainage, rectum, and head & neck) were intentionally altered to model incorrect plan delivery. Investigated errors included both systematic and random errors. Gamma analysis was conducted on both portal dose (criteria γ
resultsPredicted and acquired portal dosimetry fluences demonstrated a high agreement evident by average gamma passing rates for the clinical patient plans of 99.90%, 96.64%, and 91.87% for γ
conclusionsOwing to the high achievable spatial resolution, portal dosimetry at the Halcyon can reliably be deployed as plan-specific pre-treatment QA tool to screen for errors. It is recommended to support the fluence integrated portal dosimetry QA by independent phantom-based measurements of a random sample survey of treatment plans.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials