Evidence map›Paper›PMID 37048777›Full record

ArticleJournal of clinical medicine2023

Preclinical Application of Augmented Reality in Pediatric Craniofacial Surgery: An Accuracy Study.

Federica Ruggiero, Laura Cercenelli, Nicolas Emiliani, Giovanni Badiali, Mirko Bevini, Mino Zucchelli, Emanuela Marcelli, Achille Tarsitano

Open access · goldAbstract read
In one paragraph

Article in Journal of clinical medicine, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
10citing papers in PubMed, 1 pooled it
2.6field-weighted citation impact, top 8% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

10 citing papers in PubMed, 1 synthesis or guideline pooled it, 22 citations in OpenAlex.

  1. Pooled it
  2. Article
  3. Article
  4. Article
  5. Review
  6. Review
  7. Review
  8. Article
  9. Review
  10. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 3 institutions in 1 country.

Federica RuggieroDepartment of Biomedical and Neuromotor Science, University of Bologna, 40138 Bologna, Italy.
Laura CercenelliLaboratory of Bioengineering-eDIMES Lab, Department of Medical and Surgical Sciences (DIMEC), University of Bologna, 40138 Bologna, Italy.ORCID 0000-0001-7818-1356
Nicolas EmilianiLaboratory of Bioengineering-eDIMES Lab, Department of Medical and Surgical Sciences (DIMEC), University of Bologna, 40138 Bologna, Italy.
Giovanni BadialiDepartment of Biomedical and Neuromotor Science, University of Bologna, 40138 Bologna, Italy.ORCID 0000-0001-9833-9800
Mirko BeviniDepartment of Biomedical and Neuromotor Science, University of Bologna, 40138 Bologna, Italy.
Mino ZucchelliPediatric Neurosurgery, IRCCS Istituto delle Scienze Neurologiche di Bologna, Via Altura 3, 40138 Bologna, Italy.ORCID 0000-0002-7494-355X
Emanuela MarcelliLaboratory of Bioengineering-eDIMES Lab, Department of Medical and Surgical Sciences (DIMEC), University of Bologna, 40138 Bologna, Italy.
Achille TarsitanoDepartment of Biomedical and Neuromotor Science, University of Bologna, 40138 Bologna, Italy.ORCID 0000-0002-5815-7591
University of Bologna · ITAzienda USL di Bologna · ITIstituto delle Scienze Neurologiche di Bologna · IT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundAugmented reality (AR) allows the overlapping and integration of virtual information with the real environment. The camera of the AR device reads the object and integrates the virtual data. It has been widely applied to medical and surgical sciences in recent years and has the potential to enhance intraoperative navigation. MATERIALS AND

methodsIn this study, the authors aim to assess the accuracy of AR guidance when using the commercial HoloLens 2 head-mounted display (HMD) in pediatric craniofacial surgery. The Authors selected fronto-orbital remodeling (FOR) as the procedure to test (specifically, frontal osteotomy and nasal osteotomy were considered). Six people (three surgeons and three engineers) were recruited to perform the osteotomies on a 3D printed stereolithographic model under the guidance of AR. By means of calibrated CAD/CAM cutting guides with different grooves, the authors measured the accuracy of the osteotomies that were performed. We tested accuracy levels of ±1.5 mm, ±1 mm, and ±0.5 mm.

resultsWith the HoloLens 2, the majority of the individuals involved were able to successfully trace the trajectories of the frontal and nasal osteotomies with an accuracy level of ±1.5 mm. Additionally, 80% were able to achieve an accuracy level of ±1 mm when performing a nasal osteotomy, and 52% were able to achieve an accuracy level of ±1 mm when performing a frontal osteotomy, while 61% were able to achieve an accuracy level of ±0.5 mm when performing a nasal osteotomy, and 33% were able to achieve an accuracy level of ±0.5 mm when performing a frontal osteotomy.

conclusionsdespite this being an in vitro study, the authors reported encouraging results for the prospective use of AR on actual patients.

Indexed as

augmented realitycomputer-assisted surgerycraniofacial surgeryhead and neck surgery

Identifiers

PMID37048777
PMCPMC10095377
OpenAlexW4362617910

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked