ArticleJournal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology2023
Bionate
Article in Journal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
backgroundIntervertebral disc nucleus degeneration initiates a degenerative cascade and can induce chronic low back pain. Nucleus replacement aims to replace the nucleus while the annulus is still intact. Over time, several designs have been introduced, but the definitive solution continues to be elusive. Therefore, we aimed to create a new nucleus replacement that replicates intact intervertebral disc biomechanics, and thus has the potential for clinical applications. MATERIALS AND
methodsTwo implants with an outer ring and one (D2) with an additional midline strut were compared. Static and fatigue tests were performed with an INSTRON 8874 following the American Society for Testing and Materials F2267-04, F2346-05, 2077-03, D2990-01, and WK4863. Implant stiffness was analyzed at 0-300 N, 500-2000 N, and 2000-6000 N and implant compression at 300 N, 1000 N, 2000 N, and 6000 N. Wear tests were performed following ISO 18192-1:2008 and 18192-2:2010. GNU Octave software was used to calculate movement angles and parameters. The statistical analysis package R was used with the Deducer user interface. Statistically significant differences between the two designs were analyzed with ANOVA, followed by a post hoc analysis.
resultsD1 had better behavior in unconfined compression tests, while D2 showed a "jump." D2 deformed 1 mm more than D1. Sterilized implants were more rigid and deformed less. Both designs showed similar behavior under confined compression and when adding shear. A silicone annulus minimized differences between the designs. Wear under compression fatigue was negligible for D1 but permanent for D2. D1 suffered permanent height deformation but kept its width. D2 suffered less height loss than D1 but underwent a permanent width deformation. Both designs showed excellent responses to compression fatigue with no breaks, cracks, or delamination. At 10 million cycles, D2 showed 3-times higher wear than D1. D1 had better and more homogeneous behavior, and its wear was relatively low. It showed good mechanical endurance under dynamic loading conditions, with excellent response to axial compression fatigue loading without functional failure after long-term testing.
conclusionD1 performed better than D2. Further studies in cadaveric specimens, and eventually in a clinical setting, are recommended. Level of evidence 2c.