Evidence map›Paper›PMID 36992002›Full record

ArticleSensors (Basel, Switzerland)2023

Measures of Maximal Tactile Pressures during a Sustained Grasp Task Using a TactArray Device Have Satisfactory Reliability and Concurrent Validity in People with Stroke.

Urvashy Gopaul, Derek Laver, Leeanne Carey, Thomas Matyas, Paulette van Vliet, Robin Callister

Open access · goldAbstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.1field-weighted citation impact, top 63% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 1 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 3 institutions in 2 countries.

Urvashy GopaulKITE Research-Toronto Rehabilitation Institute, University Health Network, Toronto, ON M5G 2A2, Canada.ORCID 0000-0002-1544-5155
Derek LaverHuman Physiology, School of Biomedical Sciences & Pharmacy, College of Health, Medicine and Wellbeing, University of Newcastle, Callaghan, NSW 2308, Australia.
Leeanne CareyOccupational Therapy, School of Allied Health, Human Services and Sport, La Trobe University, Melbourne Campus, Melbourne, VIC 3086, Australia.ORCID 0000-0001-6376-8613
Thomas MatyasOccupational Therapy, School of Allied Health, Human Services and Sport, La Trobe University, Melbourne Campus, Melbourne, VIC 3086, Australia.
Paulette van VlietSchool of Health Sciences, College of Health, Medicine and Wellbeing, University of Newcastle, Newcastle, NSW 2308, Australia.
Robin CallisterHuman Physiology, School of Biomedical Sciences & Pharmacy, College of Health, Medicine and Wellbeing, University of Newcastle, Callaghan, NSW 2308, Australia.
University of Newcastle Australia · AULa Trobe University · AUUniversity Health Network · CA

Funding

University of Newcastle Australia Priority Research Centre for Stroke And Brain Injury Research Support Funding Round 2
6 · The paper itself

Abstract

Sensor-based devices can record pressure or force over time during grasping and therefore offer a more comprehensive approach to quantifying grip strength during sustained contractions. The objectives of this study were to investigate the reliability and concurrent validity of measures of maximal tactile pressures and forces during a sustained grasp task using a TactArray device in people with stroke. Participants with stroke (n = 11) performed three trials of sustained maximal grasp over 8 s. Both hands were tested in within- and between-day sessions, with and without vision. Measures of maximal tactile pressures and forces were measured for the complete (8 s) grasp duration and plateau phase (5 s). Tactile measures are reported using the highest value among three trials, the mean of two trials, and the mean of three trials. Reliability was determined using changes in mean, coefficients of variation, and intraclass correlation coefficients (ICCs). Pearson correlation coefficients were used to evaluate concurrent validity. This study found that measures of reliability assessed by changes in means were good, coefficients of variation were good to acceptable, and ICCs were very good for maximal tactile pressures using the average pressure of the mean of three trials over 8 s in the affected hand with and without vision for within-day sessions and without vision for between-day sessions. In the less affected hand, changes in mean were very good, coefficients of variations were acceptable, and ICCs were good to very good for maximal tactile pressures using the average pressure of the mean of three trials over 8 s and 5 s, respectively, in between-day sessions with and without vision. Maximal tactile pressures had moderate correlations with grip strength. The TactArray device demonstrates satisfactory reliability and concurrent validity for measures of maximal tactile pressures in people with stroke.

Indexed as

StrokeTouchHandHand StrengthHumansReproducibility of Resultsconcurrent validityfinger forcespressure sensorsreliabilitystrokesustained grasptactile

Identifiers

PMID36992002
PMCPMC10059963
OpenAlexW4328118448

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.