ArticleCVIR endovascular2023
Safety and effectiveness of bycross rotational atherectomy and aspiration device: a prospective, multi-center pre-market approval study.
Article in CVIR endovascular, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. It is linked to trial NCT03724279 (Safety and Effectiveness of ByCross Rotational and Aspiration Device for Revascularization of Total or Sub-total Occluded Peripheral Arterial Vessels of Luminal Diameter Equal to or Larger Than 3mm), which is not on this map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Safety and Effectiveness of ByCross Rotational and Aspiration Device for Revascularization of Total or Sub-total Occluded Peripheral Arterial Vessels of Luminal Diameter Equal to or Larger Than 3mm: A Prospective, Multi-center Study
Who cites it
5 citing papers in PubMed, 12 citations in OpenAlex.
- Novel technique for resolving a stuck rotational atherectomy device in a peripheral artery: a case report.American journal of cardiovascular disease · 2025Article
- Safety and Feasibility of the BYCROSSJournal of clinical medicine · 2024Article
- Innovative atherectomy device for treatment of iliac and popliteal lesions in patients with critical ischemic stage of PAOD.CVIR endovascular · 2024Article
- Revascularization with BYCROSS atherectomy device- protocol of a prospective multicenter observational study.CVIR endovascular · 2023Article
- Correction: Safety and effectiveness of bycross rotational atherectomy and aspiration device: a prospective, multi-center pre-market approval study.CVIR endovascular · 2023Article
Corrections and comments
- Erratum issued
Authors and funding
2 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo demonstrate safety and effectiveness of the novel ByCross® atherectomy system for treatment of complex femorodistal > 80% arterial stenosis. MATERIALS AND
methodsFrom September 2018 until April 2019 39 patients with 41 femorodistal lesions including the femoropopliteal and distal popliteal segments were treated in a prospective, nonrandomized pre-market approval study with 6 months follow up using the ByCross® atherectomy device (ClinicalTrials.gov identifier NCT03724279). Adjunctive treatment with balloon dilatation or stenting was allowed by the protocol. Mean patient age was 72 years with 62% male. The average lesions length was 125 ± 118 mm (30 and 450 mm) with an average reference vessel diameter of 5.2 ± 0.85 mm and a mean stenosis of 96.4 ± 6.2% based on CT or MR angiography measurements. The primary performance endpoint was defined as acute procedural success with angiographically determined residual stenosis of ≤ 50% and of ≤ 30% after atherectomy or adjunctive treatment. The primary safety endpoint was the major adverse event (MAE) rate through 30 days. Secondary endpoints were stenosis of the target lesions measured by duplex ultrasound (DUS) and the ankle-brachial pressure index (ABPI) at discharge, 30 and 180 days as well as any major adverse event (MAE) through 6 months.
resultsThe acute procedural success was achieved in 39/41 (95.12%) lesions, partially without wire guidance (11/41 (26.82%)). No embolic protection was used, and adjunctive angioplasty and stenting was performed in 40/41 (97.56%) and 12/41 (29.26%) lesions, respectively without device related MAE's at 30 days. Mean level of stenosis was 5.7% at discharge and 21.7% at 6 months FU measured by DUS. Mean ABPI was 0.8, 1.0 and 0.8 at baseline, discharge, and 6 months FU respectively.
conclusionsBased on the high technical success rate and the low rates of MAE`s through six months, the BYCROSS® Atherectomy System has shown to be safe and effective for the crossing and atherectomy of complex lower-extremity arterial occlusions.
trial registrationOctober 17, 2018,retrospectively registered. CLINICALTRIALS: gov Identifier: NCT03724279; https://clinicaltrials.gov/ct2/show/record/NCT03724279.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.