Evidence map›Paper›PMID 36979569›Full record

ReviewBiosensors2023

The Use of Sensors in Blood-Brain Barrier-on-a-Chip Devices: Current Practice and Future Directions.

András Kincses, Judit P Vigh, Dániel Petrovszki, Sándor Valkai, Anna E Kocsis, Fruzsina R Walter, Hung-Yin Lin, Jeng-Shiung Jan, Mária A Deli, András Dér

Open access · goldFull text readReview
In one paragraph

Review in Biosensors, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
2.3field-weighted citation impact, top 12% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed, 21 citations in OpenAlex.

  1. Article
  2. Review
  3. Biosensor applications in organ-on-a-chip platforms and disease modeling.Frontiers in bioengineering and biotechnology · 2026
    Review
  4. Review
  5. Review
  6. Article
  7. Review
  8. Review
  9. Review
  10. Review
  11. Article
  12. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors at 4 institutions in 2 countries.

András KincsesInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.
Judit P VighInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.
Dániel PetrovszkiInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.ORCID 0000-0002-7736-2144
Sándor ValkaiInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.ORCID 0000-0001-8479-8141
Anna E KocsisInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.
Fruzsina R WalterInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.ORCID 0000-0001-8145-2823
Hung-Yin LinDepartment of Chemical and Materials Engineering, National University of Kaohsiung, Kaohsiung 81148, Taiwan.
Jeng-Shiung JanDepartment of Chemical Engineering, National Cheng Kung University, Tainan 70101, Taiwan.ORCID 0000-0002-8379-404X
Mária A DeliInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.ORCID 0000-0001-6084-6524
András DérInstitute of Biophysics, Biological Research Centre, H-6726 Szeged, Hungary.ORCID 0000-0001-6112-884X
HUN-REN Szegedi Biológiai Kutatóközpont · HUUniversity of Szeged · HUNational Cheng Kung University · TWNational University of Kaohsiung · TW

Funding

Loránd Eötvös Research Network SA-111/2021National Research, Development and Innovation Office K-124922National Research, Development and Innovation Office NNE-129617National Research, Development and Innovation Office PD-143268National Science Technology Council, Taiwan NSTC107-2923-M-006-002-MY3
6 · The paper itself

Abstract

The application of lab-on-a-chip technologies in in vitro cell culturing swiftly resulted in improved models of human organs compared to static culture insert-based ones. These chip devices provide controlled cell culture environments to mimic physiological functions and properties. Models of the blood-brain barrier (BBB) especially profited from this advanced technological approach. The BBB represents the tightest endothelial barrier within the vasculature with high electric resistance and low passive permeability, providing a controlled interface between the circulation and the brain. The multi-cell type dynamic BBB-on-chip models are in demand in several fields as alternatives to expensive animal studies or static culture inserts methods. Their combination with integrated biosensors provides real-time and noninvasive monitoring of the integrity of the BBB and of the presence and concentration of agents contributing to the physiological and metabolic functions and pathologies. In this review, we describe built-in sensors to characterize BBB models via quasi-direct current and electrical impedance measurements, as well as the different types of biosensors for the detection of metabolites, drugs, or toxic agents. We also give an outlook on the future of the field, with potential combinations of existing methods and possible improvements of current techniques.

Indexed as

Blood-Brain BarrierBrainAnimalsBiological TransportCell Culture TechniquesHumansLab-On-A-Chip Devicesbiosensorblood-brain barriercell surface chargechemosensorelectrical impedance spectroscopyoptical sensororgan-on-a-chipstreaming potentialtransendothelial electrical resistance

Identifiers

PMID36979569
PMCPMC10046513
OpenAlexW4323565898

What OpenQuestion holds

Textfull text, public
LicenceCC BY
measurements read2
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.