Evidence map›Paper›PMID 36904759›Full record

ArticleSensors (Basel, Switzerland)2023

Using a Smartwatch to Record Precordial Electrocardiograms: A Validation Study.

Joske van der Zande, Marc Strik, Rémi Dubois, Sylvain Ploux, Saer Abu Alrub, Théo Caillol, Mathieu Nasarre, Dirk W Donker, Eline Oppersma, Pierre Bordachar

Full text read
In one paragraph

Article in Sensors (Basel, Switzerland), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 13 papers.

0numbers the graph read from it
0cells of the map it votes in
13citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

13 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Wearable-Echo-FM: An ECG-echo foundation model for single lead electrocardiography.medRxiv : the preprint server for health sciences · 2025
    Article
  6. Review
  7. Article
  8. Review
  9. Article
  10. Review
  11. Article
  12. Article
  13. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Joske van der ZandeIHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, F-33600 Pessac, France.ORCID 0000-0002-4998-6363
Marc StrikIHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, F-33600 Pessac, France.
Rémi DuboisIHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, F-33600 Pessac, France.ORCID 0000-0002-3320-6086
Sylvain PlouxIHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, F-33600 Pessac, France.
Saer Abu AlrubCardio-Thoracic Unit, Bordeaux University Hospital (CHU), F-33600 Pessac, France.ORCID 0000-0002-6638-4233
Théo CaillolIHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, F-33600 Pessac, France.
Mathieu NasarreCardio-Thoracic Unit, Bordeaux University Hospital (CHU), F-33600 Pessac, France.
Dirk W DonkerCardiovascular and Respiratory Physiology, TechMed Centre, University of Twente, 7522 NB Enschede, The Netherlands.ORCID 0000-0001-6496-2768
Eline OppersmaCardiovascular and Respiratory Physiology, TechMed Centre, University of Twente, 7522 NB Enschede, The Netherlands.ORCID 0000-0002-0150-306X
Pierre BordacharIHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, F-33600 Pessac, France.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Smartwatches that support the recording of a single-lead electrocardiogram (ECG) are increasingly being used beyond the wrist, by placement on the ankle and on the chest. However, the reliability of frontal and precordial ECGs other than lead I is unknown. This clinical validation study assessed the reliability of an Apple Watch (AW) to obtain conventional frontal and precordial leads as compared to standard 12-lead ECGs in both subjects without known cardiac anomalies and patients with underlying heart disease. In 200 subjects (67% with ECG anomalies), a standard 12-lead ECG was performed, followed by AW recordings of the standard Einthoven leads (leads I, II, and III) and precordial leads V1, V3, and V6. Seven parameters (P, QRS, ST, and T-wave amplitudes, PR, QRS, and QT intervals) were compared through a Bland-Altman analysis, including the bias, absolute offset, and 95% limits of agreement. AW-ECGs recorded on the wrist but also beyond the wrist had similar durations and amplitudes compared to standard 12-lead ECGs. Significantly greater amplitudes were measured by the AW for R-waves in precordial leads V1, V3, and V6 (+0.094 mV, +0.149 mV, +0.129 mV, respectively, all

Indexed as

ElectrocardiographyHeart DiseasesArrhythmias, CardiacHumansReproducibility of ResultsThoraxambulatory electrocardiographyApple Watchcardiac arrhythmiasmobile healthreliabilitywearables

Identifiers

PMID36904759
PMCPMC10007514

What OpenQuestion holds

Textfull text, public
LicenceCC BY
measurements read11
Read underepoch 390

Registered trials

None linked