Performance evaluation of SimPET-L and SimPET-XL: MRI-compatible small-animal PET systems with rat-body imaging capability.

Minjee Seo, Guen Bae Ko, Kyeong Yun Kim, Jeong-Whan Son, Jung Woo Byun, Yun-Sang Lee, Kyeong Min Kim, Jang Woo Park, Kipom Kim, Taekwan Lee and 1 more

Abstract read
In one paragraph

Article in EJNMMI physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Minjee SeoDepartment of Biomedical Sciences, Seoul National University College of Medicine, Seoul, 03080, South Korea.
Guen Bae KoBrightonix Imaging Inc., Seoul, 04782, South Korea.
Kyeong Yun KimBrightonix Imaging Inc., Seoul, 04782, South Korea.
Jeong-Whan SonBrightonix Imaging Inc., Seoul, 04782, South Korea.
Jung Woo ByunDepartment of Nuclear Medicine, Seoul National University College of Medicine, 103 Daehak-ro, Jongno-gu, Seoul, 03080, South Korea.
Yun-Sang LeeDepartment of Nuclear Medicine, Seoul National University College of Medicine, 103 Daehak-ro, Jongno-gu, Seoul, 03080, South Korea.
Kyeong Min KimKorea Institute of Radiological and Medical Sciences, Seoul, 01812, South Korea.
Jang Woo ParkKorea Institute of Radiological and Medical Sciences, Seoul, 01812, South Korea.
Kipom KimBrain Research Core Facility, Korea Brain Research Institute, 61 Cheomdan-ro, Dong-gu, Daegu, 41062, South Korea.
Taekwan LeeBrain Research Core Facility, Korea Brain Research Institute, 61 Cheomdan-ro, Dong-gu, Daegu, 41062, South Korea. tklee@kbri.re.kr.
Jae Sung LeeDepartment of Biomedical Sciences, Seoul National University College of Medicine, Seoul, 03080, South Korea. jaes@snu.ac.kr.ORCID http://orcid.org/0000-0001-7623-053X

Funding

Korea Brain Research Institute 22-BR-05-02Korea Medical Device Development Fund 1711137868
6 · The paper itself

Abstract

backgroundSimPET-L and SimPET-XL have recently been introduced with increased transaxial fields of view (FOV) compared with their predecessors (SimPET™ and SimPET-X), enabling whole-body positron emission tomography (PET) imaging of rats. We conducted performance evaluations of SimPET-L and SimPET-XL and rat-body imaging with SimPET-XL to demonstrate the benefits of increased axial and transaxial FOVs. PROCEDURES: The detector blocks in SimPET-L and SimPET-XL consist of two 4 × 4 silicon photomultiplier arrays coupled with 20 × 9 array lutetium oxyorthosilicate crystals. SimPET-L and SimPET-XL have an inner diameter (bore size) of 7.6 cm, and they are composed of 40 and 80 detector blocks yielding axial lengths of 5.5 and 11 cm, respectively. Each system was evaluated according to the National Electrical Manufacturers Association NU4-2008 protocol. Rat imaging studies, such as

resultsThe radial resolutions at the axial center measured using the filtered back projection, 3D ordered-subset expectation maximization (OSEM), and 3D OSEM with point spread functions correction were 1.7, 0.82, and 0.82 mm FWHM in SimPET-L and 1.7, 0.91, and 0.91 mm FWHM in SimPET-XL, respectively. The peak sensitivities of SimPET-L and SimPET-XL were 6.30% and 10.4% for an energy window of 100-900 keV and 4.44% and 7.25% for a window of 250-750 keV, respectively. The peak noise equivalent count rate with an energy window of 250-750 keV was 249 kcps at 44.9 MBq for SimPET-L and 349 kcps at 31.3 MBq for SimPET-XL. In SimPET-L, the uniformity was 4.43%, and the spill-over ratios in air- and water-filled chambers were 5.54% and 4.10%, respectively. In SimPET-XL, the uniformity was 3.89%, and the spill-over ratio in the air- and water-filled chambers were 3.56% and 3.60%. Moreover, SimPET-XL provided high-quality images of rats.

conclusionSimPET-L and SimPET-XL show adequate performance compared with other SimPET systems. In addition, their large transaxial and long axial FOVs provide imaging capability for rats with high image quality.

Indexed as

InstrumentationMolecular imagingNEMA performancePET/MRI