Calibration procedure and biomechanical validation of an universal six degree-of-freedom robotic system for hip joint testing.

Michal Rychlik, Georg Wendland, Michal Jackowski, Roland Rennert, Klaus-Dieter Schaser, Joerg Nowotny

Open access · goldAbstract read
In one paragraph

Article in Journal of orthopaedic surgery and research, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
2.5field-weighted citation impact, top 11% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 8 citations in OpenAlex.

  1. Article
  2. Article
  3. Exploration of the Advanced VIVOBioengineering (Basel, Switzerland) · 2024
    Article
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 4 institutions in 2 countries.

Michal RychlikUniversity Center of Orthopaedics, Trauma and Plastic Surgery, University Hospital Carl Gustav Carus, TU Dresden, Fetscherstraße 74, 01307, Dresden, Germany.
Georg WendlandIMA Materialforschung Und Anwendungstechnik GmbH, Dresden, Germany.
Michal JackowskiIMA Materialforschung Und Anwendungstechnik GmbH, Dresden, Germany.
Roland RennertIMA Materialforschung Und Anwendungstechnik GmbH, Dresden, Germany.
Klaus-Dieter SchaserUniversity Center of Orthopaedics, Trauma and Plastic Surgery, University Hospital Carl Gustav Carus, TU Dresden, Fetscherstraße 74, 01307, Dresden, Germany.
Joerg NowotnyUniversity Center of Orthopaedics, Trauma and Plastic Surgery, University Hospital Carl Gustav Carus, TU Dresden, Fetscherstraße 74, 01307, Dresden, Germany. joerg.nowotny@uniklinikum-dresden.de.
IMA Materialforschung und Anwendungstechnik · DEKlinik und Poliklinik für Psychotherapie und Psychosomatik · DETechnische Universität Dresden · DEUniversity Hospital Carl Gustav Carus · DE

Funding

Sächsische Aufbaubank project No. 205.1009
6 · The paper itself

Abstract

purposeAmong various test methods for different human joints, the use of robot systems has attracted major interest and inherits the potential to become a gold standard in biomechanical testing in the future. A key issue associated with those robot-based platforms is the accurate definition of parameters, e.g., tool center point (TCP), length of tool or anatomical trajectories of movements. These must be precisely correlated to the physiological parameters of the examined joint and its corresponding bones. Exemplified for the human hip joint, we are creating an accurate calibration procedure for a universal testing platform by using a six degree-of-freedom (6 DOF) robot and optical tracking system for recognition of anatomical movements of the bone samples.

methodsA six degree-of-freedom robot (TX 200, Stäubli) has been installed and configured. The physiological range of motion of the hip joint composed of a femur and a hemipelvis was recorded with an optical 3D movement and deformation analysis system (ARAMIS, GOM GmbH). The recorded measurements were processed by automatic transformation procedure (created in Delphi software) and evaluated in 3D CAD system.

resultsThe physiological ranges of motion were reproduced for all degrees of freedom with the six degree-of-freedom robot in adequate accuracy. With the establishment of a special calibration procedure by using a combination of different coordinate systems, we were able to achieve a standard deviation of the TCP depending of the axis between 0.3 and 0.9 mm and for the length of tool between + 0.67 and - 0.40 mm (3D CAD processing) resp. + 0.72 mm to - 0.13 mm (Delphi transformation). The accuracy between the manual and robotic movement of the hip shows an average deviation between - 0.36 and + 3.44 mm for the points on the movement trajectories.

conclusionA six degree-of-freedom robot is appropriate to reproduce the physiological range of motion of the hip joint. The described calibration procedure is universal and can be used for hip joint biomechanical tests allowing to apply clinically relevant forces and investigate testing stability of reconstructive osteosynthesis implant/endoprosthetic fixations, regardless of the length of the femur, size of the femoral head and acetabulum or whether the entire pelvis or only the hemipelvis will be used.

Indexed as

Robotic Surgical ProceduresAcetabulumCalibrationFemurHip JointHumansBiomechanicsHip jointKinematicsSix degree-of-freedom robotTesting

Identifiers

PMID36869379
PMCPMC9983254
OpenAlexW4323043852

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.