ArticleJournal of applied clinical medical physics2023

Impact of electronic portal image quality on Elekta AQUA

Craig A J Norvill

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Impact of electronic portal image quality on Elekta AQUAJournal of applied clinical medical physics · 2023
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

1 author.

Craig A J NorvillGenesis Care, Perth, Australia.ORCID https://orcid.org/0000-0002-7724-5335

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe collimator radiation isocenter position determined in AQUA

methodsThe AQUA test determines radiation isocenter using the mean field center of nine 4 cm × 4 cm electronic portal imager (EPID) exposures at equidistant collimator angles. Impact of EPID image quality on AQUA reported isocenter for thirteen Elekta linear accelerators with Agility MLC heads were evaluated.

resultsOf the thirteen, three had visually and quantitatively identifiable artifacts. For the ten good EPID's there was a systematic 0.25 mm offset of the MATLAB calculated mean field center relative to AQUA in the X-axis and Y-axis. This corresponds to one image pixel and was found to be due to differences in software co-ordinate convention. After subtracting this offset there was no significant difference in AQUA and MATLAB calculated isocenter.

conclusionsFor the three machines with poor image quality there was a demonstrated variation in AQUA calculated field center and therefore radiation isocenter relative to MATLAB. Restricting the region of interest (ROI) in AQUA software to only the irradiated section of the EPID brought AQUA and MATLAB result for these three machines into agreement.

Indexed as

ElectronicsParticle AcceleratorsHumansSoftwareSwedenAQUAEPIDIsocenter

Identifiers

PMID36855933
PMCPMC10113691

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.