Evidence map›Paper›PMID 36844740›Full record

ArticleFrontiers in cardiovascular medicine2023

Wrist-worn device combining PPG and ECG can be reliably used for atrial fibrillation detection in an outpatient setting.

Harri Juhani Saarinen, Atte Joutsen, Kirsi Korpi, Tuomas Halkola, Marko Nurmi, Jussi Hernesniemi, Antti Vehkaoja

Abstract read
In one paragraph

Article in Frontiers in cardiovascular medicine, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed.

  1. Review
  2. Review
  3. Review
  4. Article
  5. Review
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Harri Juhani SaarinenHeart Hospital, Tampere University Hospital, Tampere, Finland.
Atte JoutsenFaculty of Medicine and Health Technology, Tampere University, Tampere, Finland.
Kirsi KorpiHeart Hospital, Tampere University Hospital, Tampere, Finland.
Tuomas HalkolaPulseOn Oy, Espoo, Finland.
Marko NurmiPulseOn Oy, Espoo, Finland.
Jussi HernesniemiHeart Hospital, Tampere University Hospital, Tampere, Finland.
Antti VehkaojaFaculty of Medicine and Health Technology, Tampere University, Tampere, Finland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Aims: The aim was to validate the performance of a monitoring system consisting of a wrist-worn device and a data management cloud service intended to be used by medical professionals in detecting atrial fibrillation (AF). Methods: Thirty adult patients diagnosed with AF alone or AF with concomitant flutter were recruited. Continuous photoplethysmogram (PPG) and intermittent 30 s Lead I electrocardiogram (ECG) recordings were collected over 48 h. The ECG was measured four times a day at prescheduled times, when notified due to irregular rhythm detected by PPG, and when self-initiated based on symptoms. Three-channel Holter ECG was used as the reference. Results: The subjects recorded a total of 1,415 h of continuous PPG data and 3.8 h of intermittent ECG data over the study period. The PPG data were analyzed by the system's algorithm in 5-min segments. The segments containing adequate amounts, at least ~30 s, of adequate quality PPG data for rhythm assessment algorithm, were included. After rejecting 46% of the 5-min segments, the remaining data were compared with annotated Holter ECG yielding AF detection sensitivity and specificity of 95.6 and 99.2%, respectively. The ECG analysis algorithm labeled 10% of the 30-s ECG records as inadequate quality and these were excluded from the analysis. The ECG AF detection sensitivity and specificity were 97.7 and 89.8%, respectively. The usability of the system was found to be good by both the study subjects and the participating cardiologists. Conclusion: The system comprising of a wrist device and a data management service was validated to be suitable for use in patient monitoring and in the detection of AF in an ambulatory setting.

Indexed as

ambulatoryatrial fibrillationelectrocardio graphyphotoplethysmographywearable sensors

Identifiers

PMID36844740
PMCPMC9949528

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.