Evidence map›Paper›PMID 36835888›Full record

ArticleJournal of clinical medicine2023

Artificial Intelligence (AI) versus POCUS Expert: A Validation Study of Three Automatic AI-Based, Real-Time, Hemodynamic Echocardiographic Assessment Tools.

Eyal Gohar, Amit Herling, Mor Mazuz, Gal Tsaban, Tomer Gat, Sergio Kobal, Lior Fuchs

Open access · goldAbstract read
In one paragraph

Article in Journal of clinical medicine, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 28 papers.

0numbers the graph read from it
0cells of the map it votes in
28citing papers in PubMed
11.1field-weighted citation impact, top 1% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

28 citing papers in PubMed, 52 citations in OpenAlex.

  1. Trial
  2. Article
  3. Review
  4. Article
  5. Review
  6. Article
  7. Article
  8. AI-Enhanced POCUS in Emergency Care.Diagnostics (Basel, Switzerland) · 2026
    Review
  9. Review
  10. Article
  11. Review
  12. Review
  13. Article
  14. Review
  15. Review
  16. Review
  17. Article
  18. Review
  19. Review
  20. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Eyal GoharBen-Gurion University of the Negev, Beer Sheva 8410501, Israel.ORCID 0000-0002-2371-065X
Amit HerlingBen-Gurion University of the Negev, Beer Sheva 8410501, Israel.
Mor MazuzBen-Gurion University of the Negev, Beer Sheva 8410501, Israel.
Gal TsabanSoroka Medical Center, Beer Sheva 84101, Israel.ORCID 0000-0003-4234-6504
Tomer GatSoroka Medical Center, Beer Sheva 84101, Israel.ORCID 0000-0002-0573-8376
Sergio KobalSoroka Medical Center, Beer Sheva 84101, Israel.ORCID 0000-0002-5167-0448
Lior FuchsSoroka Medical Center, Beer Sheva 84101, Israel.
Soroka Medical Center · ILBen-Gurion University of the Negev · IL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundPoint Of Care Ultra-Sound (POCUS) is an operator dependent modality. POCUS examinations usually include 'Eyeballing' the inspected anatomical structure without conducting accurate measurements due to complexity and insufficient time. Automatic real time measuring tools can make accurate measurements fast and simple and dramatically increase examination reliability while saving the operator much time and effort. In this study we aim to assess three automatic tools which are integrated into the Venue™ device by GE: the automatic ejection fraction, velocity time integral, and inferior vena cava tools in comparison to the gold standard-an examination by a POCUS expert.

methodsA separate study was conducted for each of the three automatic tools. In each study, cardiac views were acquired by a POCUS expert. Relevant measurements were taken by both an auto tool and a POCUS expert who was blinded to the auto tool's measurement. The agreement between the POCUS expert and the auto tool was measured for both the measurements and the image quality using a Cohen's Kappa test.

resultsAll three tools have shown good agreement with the POCUS expert for high quality views: auto LVEF (0.498;

conclusionsThe Venue™ show a high agreement with a POCUS expert for high quality views. This shows that auto tools can provide reliable real time assistance in performing accurate measurements, but do not reduce the need of a good image acquisition technique.

Indexed as

Artificial IntelligenceBen Gurion University of the NegevEchocardiographyEjection Fraction (EF)Inferior Vena Cava measurementIntensive Care Unit (ICU)IVC collapsibility index (IVC CI)Point Of Care UltraSound (POCUS)Real Time US toolSoroka University Medical CenterVelocity Time Integral (VTI)

Identifiers

PMID36835888
PMCPMC9959768
OpenAlexW4319596896

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.