ArticleAnnals of biomedical engineering2023

Experimental Protocol and Phantom Design and Development for Performance Characterization of Conventional Devices for Peripheral Vascular Interventions.

Yara Alawneh, James J Zhou, Alykhan Sewani, Mohammadmahdi Tahmasebi, Trisha L Roy, Ahmed Kayssi, Andrew D Dueck, Graham A Wright, M Ali Tavallaei

Abstract read
PubMed Publisher
In one paragraph

Article in Annals of biomedical engineering, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Yara AlawnehFaculty of Engineering and Architectural Science, Toronto Metropolitan University (Formerly Ryerson University), Toronto, ON, Canada.ORCID http://orcid.org/0000-0002-2146-9575
James J ZhouFaculty of Engineering and Architectural Science, Toronto Metropolitan University (Formerly Ryerson University), Toronto, ON, Canada.ORCID http://orcid.org/0000-0002-9755-897X
Alykhan SewaniFaculty of Engineering and Architectural Science, Toronto Metropolitan University (Formerly Ryerson University), Toronto, ON, Canada.ORCID http://orcid.org/0000-0001-8037-5184
Mohammadmahdi TahmasebiFaculty of Engineering and Architectural Science, Toronto Metropolitan University (Formerly Ryerson University), Toronto, ON, Canada.
Trisha L RoyHouston Methodist DeBakey Heart & Vascular Center, Houston Methodist Hospital, Houston, TX, USA.ORCID http://orcid.org/0000-0002-2180-1275
Ahmed KayssiDepartment of Vascular Surgery, Sunnybrook Health Sciences Centre, Toronto, ON, Canada.
Andrew D DueckDepartment of Vascular Surgery, Sunnybrook Health Sciences Centre, Toronto, ON, Canada.
Graham A WrightUniversity of Toronto, Toronto, ON, Canada.
M Ali TavallaeiFaculty of Engineering and Architectural Science, Toronto Metropolitan University (Formerly Ryerson University), Toronto, ON, Canada. ali.tavallaei@torontomu.ca.ORCID http://orcid.org/0000-0002-8057-8536

Funding

Natural Sciences and Engineering Research Council of Canada NSERC DG
6 · The paper itself

Abstract

Conventional catheter-based interventions for treating peripheral artery disease suffer high failure and complication rates. The mechanical interactions with the anatomy constrain catheter controllability, while their length and flexibility limit their pushability. Also, the 2D X-ray fluoroscopy guiding these procedures fails to provide sufficient feedback about the device location relative to the anatomy. Our study aims to quantify the performance of conventional non-steerable (NS) and steerable (S) catheters in phantom and ex vivo experiments. In a 10 mm diameter, 30 cm long artery phantom model, with four operators, we evaluated the success rate and crossing time in accessing 1.25 mm target channels, the accessible workspace, and the force delivered through each catheter. For clinical relevance, we evaluated the success rate and crossing time in crossing ex vivo chronic total occlusions. For the S and NS catheters, respectively, users successfully accessed 69 and 31% of the targets, 68 and 45% of the cross-sectional area, and could deliver 14.2 and 10.2 g of mean force. Using a NS catheter, users crossed 0.0 and 9.5% of the fixed and fresh lesions, respectively. Overall, we quantified the limitations of conventional catheters (navigation, reachable workspace, and pushability) for peripheral interventions; this can serve as a basis for comparison with other devices.