Container closure integrity testing and process validation of closed system transfer devices for aseptic reconstitution of drug vials connected to fluid bags.

Roland B van den Berg, Kübra Akgöl, Eleonora L Swart, Bastiaan Nuijen, Mirjam Crul

Open access · greenAbstract readValidation Study
In one paragraph

Article in European journal of hospital pharmacy : science and practice, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.8field-weighted citation impact, top 33% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 1 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 1 country.

Roland B van den BergDepartment of Pharmacy and Clinical Pharmacology, Amsterdam UMC, Vrije Universiteit Amsterdam, Amsterdam, The Netherlands r.vandenberg2@amsterdamumc.nl.ORCID http://orcid.org/0000-0002-2285-781X
Kübra AkgölDepartment of Pharmacy and Pharmacology, Netherlands Cancer Institute, Antoni van Leeuwenhoek, Amsterdam, The Netherlands.
Eleonora L SwartDepartment of Pharmacy and Clinical Pharmacology, Amsterdam UMC, Vrije Universiteit Amsterdam, Amsterdam, The Netherlands.
Bastiaan NuijenDepartment of Pharmacy and Pharmacology, Netherlands Cancer Institute, Antoni van Leeuwenhoek, Amsterdam, The Netherlands.
Mirjam CrulDepartment of Pharmacy and Clinical Pharmacology, Amsterdam UMC, Vrije Universiteit Amsterdam, Amsterdam, The Netherlands.
Amsterdam University Medical Centers · NLThe Netherlands Cancer Institute · NL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesThe closure integrity and process validation of closed system transfer devices (CSTDs) should be confirmed before implementation in clinical settings. We aimed to investigate the closure integrity and validate the aseptic procedure of two types of CSTDs by using a combination of the dye ingress test and a media fill test.

methodsThe dye ingress test with methylene blue was used for both CSTDs with 10 samples of drug vials of three brands. A media fill test was performed with both CSTDs (300 samples per CSTD, 150 carried out in a safety cabinet and 150 under non-classified environmental conditions).

resultsIn all samples of both CSTDs, methylene blue was absent after visual inspection and spectrophotometric analysis. The nutrient media of one sample with CSTD A and none of the CSTD B samples were contaminated when reconstituted in a GMP grade A environment. Under non-classified environmental conditions, one sample of CSTD A and two samples of CSTD B were contaminated.

conclusionsBoth CSTDs connected to the drug vials met the terms of closure integrity by using the dye ingress test. The aseptic procedure of CSTD B was validated with the media fill test when reconstituted in a GMP grade A environment, but failed for CSTD A. Both CSTDs failed the media fill test when reconstituted under non-classified environmental conditions.

Indexed as

Drug PackagingMethylene BlueAsepsisColoring AgentsDrug ContaminationHumansColoring AgentsMethylene Bluechemical safetydrug incompatibilitypharmaceutical preparationspharmacopoeiasafety

Identifiers

PMID36792349
PMCPMC11265557
OpenAlexW4320912324

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.