ArticleEuropean radiology experimental2023
Visual and quantitative assessment of hip implant-related metal artifacts at low field MRI: a phantom study comparing a 0.55-T system with 1.5-T and 3-T systems.
Article in European radiology experimental, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed.
- Assessment of MR-Induced Heating Effects in Four Common Veterinary Implants at 1.5 and 3.0 T: A Phantom Study.Veterinary radiology & ultrasound : the official journal of the American College of Veterinary Radiology and the International Veterinary Radiology Association · 2026Article
- Net zero emission MR imaging using a permanent 0.4 T magnet.Magma (New York, N.Y.) · 2026Article
- Open-Source Simulator of Imaging Near Metal at Arbitrary Magnetic Field Strengths.Magnetic resonance in medicine · 2026Article
- Intraoral Coil Arrays for Single-Tooth Dental MRI.Magnetic resonance in medicine · 2026Article
- Acceleration of slice encoding for metal artifact correction at 0.55 T using hexagonal sampling.Magnetic resonance in medicine · 2026Article
- Low-field and portable MRI technology: advancements and innovations.European radiology experimental · 2025Review
- Imaging near titanium total hip arthroplasty at 0.55 T compared with 3 T.Magnetic resonance in medicine · 2025Article
- Improved metal suppression using new generation low-field MRI: a biophantom feasibility study.Skeletal radiology · 2025Article
- Clinical MR imaging of patients with spinal hardware at 0.55T: comparison of diagnostic assessment and metal artifact appearance with 1.5T.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2025Article
- Influence of ear tags on the results of body composition analysis in mice.Animal models and experimental medicine · 2024Article
- The potential of low-field MRI in abdominal imaging.European radiology · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundTo investigate hip implant-related metal artifacts on a 0.55-T system compared with 1.5-T and 3-T systems.
methodsTotal hip arthroplasty made of three different alloys were evaluated in a water phantom at 0.55, 1.5, and 3 T using routine protocols. Visually assessment (VA) was performed by three readers using a Likert scale from 0 (no artifacts) to 6 (extremely severe artifacts). Quantitative assessment (QA) was performed using the coefficient of variation (CoV) and the fraction of voxels within a threshold of the mean signal intensity compared to an automatically defined region of interest (FVwT). Agreement was evaluated using intra/inter-class correlation coefficient (ICC).
resultsInterreader agreement of VA was strong-to-moderate (ICC 0.74-0.82). At all field strengths (0.55-T/1.5-T/3-T), artifacts were assigned a lower score for titanium (Ti) alloys (2.44/2.9/2.7) than for stainless steel (Fe-Cr) (4.1/3.9/5.1) and cobalt-chromium (Co-Cr) alloys (4.1/4.1/5.2) (p < 0.001 for both). Artifacts were lower for 0.55-T and 1.5-T than for 3-T systems, for all implants (p ≤ 0.049). A strong VA-to-QA correlation was found (r = 0.81; p < 0.001); CoV was lower for Ti alloys than for Fe-Cr and Co-Cr alloys at all field strengths. The FVwT showed a negative correlation with VA (-0.68 < r < -0.84; p < 0.001).
conclusionsArtifact intensity was lowest for Ti alloys at 0.55 T. For other alloys, it was similar at 0.55 T and 1.5 T, higher at 3 T. Despite an inferior gradient system and a larger bore width, the 0.55-T system showed the same artifact intensity of the 1.5-T system.