Evidence map›Paper›PMID 36738388›Full record

ArticleJournal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing2023

Prospective study of zero-fluoroscopy laser balloon pulmonary vein isolation for the management of atrial fibrillation.

Jamario Skeete, Henry D Huang, David Kenigsberg

Abstract read
PubMed Publisher
In one paragraph

Article in Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed, 1 pooled it
0.4field-weighted citation impact, top 36% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 1 synthesis or guideline pooled it, 2 citations in OpenAlex.

  1. Safety and efficacy of intracardiac echocardiography in atrial fibrillation ablation: a meta-analysis.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
    Pooled it
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 2 institutions in 1 country.

Jamario SkeeteSection of Electrophysiology, Division of Cardiology, Rush University Medical Center, 1717 W, Congress Parkway, Kellogg, Suite 320, Chicago, IL, 60612, USA. jamario_r_skeete@rush.edu.ORCID http://orcid.org/0000-0003-3303-7101
Henry D HuangSection of Electrophysiology, Division of Cardiology, Rush University Medical Center, 1717 W, Congress Parkway, Kellogg, Suite 320, Chicago, IL, 60612, USA.ORCID http://orcid.org/0000-0002-4103-4317
David KenigsbergCardiac Electrophysiology, Florida Heart Rhythm Specialists, Fort Lauderdale, FL, USA.ORCID http://orcid.org/0000-0001-9549-4004
Rush University Medical Center · USHeart Rhythm Society · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundIn recent years, there has been increased focus on the development of safe and effective strategies to minimize and ultimately eliminate fluoroscopy use in the electrophysiology lab due to the inherent risks to patients and staff associated with this imaging source. However, studies examining these innovative fluoroless strategies for pulmonary vein isolation (PVI) using catheters without direct 3D mapping system integration are lacking. We sought to develop a method to perform zero-fluoroscopy laser balloon PVI for patients with atrial fibrillation (AF), and to test the safety and efficacy of this approach.

methodsWe developed a standardized method for performing PVI using the X3 laser balloon (LB) system, 3-dimensional electroanatomic mapping (3D-EAM) and intracardiac echocardiography (ICE) in a cohort of patients with symptomatic AF. The primary endpoint of the study was the ability to perform PVI without the use of fluoroscopy. Secondary outcomes were rate of successful transseptal puncture on first attempt, first pass isolation of target PVs, mean procedural time, active laser time to achieve PVI, need for use of supplemental energy sources, and procedural complication rates.

resultsTwo hundred consecutive patients undergoing PVI were recruited in the study. In the zero-fluoroscopy group, LB PVI was successfully performed in 100% of participants (n = 100) without the need for fluoroscopy. Transseptal access was achieved in 100% of cases on the first attempt. Successful first pass PVI was achieved in 360 of the 387 pulmonary veins attempted (93%). Mean procedural time was 68.2 ± 16.2 min in the zero-fluoroscopy group versus 67.5 ± 17.0 min in the conventional fluoroscopy group. PVI was able to be achieved in 100% of cases in both groups without need for use of supplemental energy sources. In the zero-fluoroscopy group there were minimal complications, with 3% of all cases having groin complications and 1 patient with a pericardial effusion noted post-procedure which was managed conservatively.

conclusionsWe demonstrated that successful zero-fluoroscopy LB PVI could be performed at a single high-volume center by experienced operators in an effective manner, without significant complications.

Indexed as

Atrial FibrillationCatheter AblationPulmonary VeinsFluoroscopyHumansLasersProspective StudiesTreatment OutcomeAtrial fibrillation ablationFluorolessLaser balloon ablationZero-fluoroscopy

Identifiers

PMID36738388
OpenAlexW4319216357

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked