Evidence map›Paper›PMID 36704641›Full record

ArticleJADA foundational science2022

Effects of abutment screw preload and preload simulation techniques on dental implant lifetime.

Megha Satpathy, Rose M Jose, Yuanyuan Duan, Jason A Griggs

Erratum issuedAbstract read
In one paragraph

Article in JADA foundational science, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 14 papers.

0numbers the graph read from it
0cells of the map it votes in
14citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

14 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. [Impact of ultrasonic scaling on abutment screw torque in implant restorations].Beijing da xue xue bao. Yi xue ban = Journal of Peking University. Health sciences · 2026
    Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Screening dental implant design parameters for effect on the fatigue limit of reduced-diameter implants.Dental materials : official publication of the Academy of Dental Materials · 2025
    Article
  11. Article
  12. Structural and torque changes in implant components of different diameters subjected to mechanical fatigue.Dental materials : official publication of the Academy of Dental Materials · 2024
    Article
  13. Article
  14. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

4 authors.

Megha SatpathyBiomedical Materials Science, University of Mississippi Medical Center, Jackson, MS.
Rose M JoseBiomedical Materials Science, University of Mississippi Medical Center, Jackson, MS.
Yuanyuan DuanBiomedical Materials Science, University of Mississippi Medical Center, Jackson, MS.
Jason A GriggsBiomedical Materials Science, University of Mississippi Medical Center, Jackson, MS.

Funding

Design Optimization of Reduced-Diameter Implants in Simulated and Cadaver BoneR01DE026144 · NIDCR · UNIVERSITY OF MISSISSIPPI MED CTR · PI GRIGGS, JASON A · 2019 to 2023
$1.8M
NIDCR NIH HHS R01 DE026144
6 · The paper itself

Abstract

Background: This study aimed to investigate how the predicted implant fatigue lifetime is affected by the loss of connector screw preload and the finite element analysis method used to simulate preload. Methods: A dental implant assembly (DI1, Biomet-3i external hex; Zimmer Biomet) was scanned using microcomputed tomography and measured using Mimics software (Materialise) and an optical microscope. Digital replicas were constructed using SolidWorks software (Dassault Systèmes). The material properties were assigned in Abaqus (Dassault Systèmes). An external load was applied at 30° off-axial loading. Eight levels of connector screw preload (range, 0-32 Ncm) were simulated for DI1. This assembly and an additional model (DI2) having a longer and narrower screw were compared regarding their fatigue limits (using fe-safe software [Dassault Systèmes]) for 2 preloading methods: (1) adding preload torque or (2) adding bolt axial tension. Results: The maximum von Mises stresses of DI1 (on the connector screw threads) with and without preload were 439.90 MPa and 587.90 MPa. The predicted fatigue limit was the same for preloads from 100% through 80% of the manufacturer Conclusions: A substantial loss of preload can be accommodated without compromising the fatigue resistance of this dental implant. Computer models should be constructed using torque instead of a bolt axial tension.

Indexed as

failurefatiguefinite element analysisfrictionImplantsmechanical testing

Identifiers

PMID36704641
PMCPMC9873498

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked