Dose Agreement Analysis of Treatment Planning System-Calculated Doses and Markus Chamber-Measured Doses in the Near-Surface Region for Breast Cancer Patients' Conformal Treatment Plans.

Garima Gaur, Raja Paramjeet Singh Banipal, Pardeep Garg, Om Prakash Gurjar, Gurpreet Kaur, Sheetal Sachdeva, Romikant Grover, Manraj Singh Kang, Vinod Kumar Dangwal

Open access · diamondAbstract read
In one paragraph

Article in Journal of medical physics. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.5field-weighted citation impact, top 33% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 3 citations in OpenAlex.

  1. Multifunctional Fluorescent Microgel-Embedded Hydrogels for Temperature and X-Ray Sensing.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2025
    Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 3 institutions in 1 country.

Garima GaurDepartment of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
Raja Paramjeet Singh BanipalDepartment of Radiation Oncology, Government Medical College, Patiala, Punjab, India.
Pardeep GargDepartment of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
Om Prakash GurjarDepartment of Radiotherapy, Government Cancer Hospital, MGM Medical College, Indore, Madhya Pradesh, India.
Gurpreet KaurDepartment of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
Sheetal SachdevaAdvance Cancer Centre, Bhatinda, Punjab, India.
Romikant GroverDepartment of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
Manraj Singh KangDepartment of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
Vinod Kumar DangwalDepartment of Radiation Oncology, Government Medical College, Patiala, Punjab, India.
Guru Gobind Singh Medical College and Hospital · INGovernment Medical College · INMahatma Gandhi Memorial Medical College · IN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Surface/skin dose measurement is one of the most challenging tasks for clinical dosimetry in radiotherapy and comparison with almost all the commercially available treatment planning systems (TPSs) brings a significant variation with the measured dose. Aims and Objectives: In the current study, doses calculated from the TPS in the near-surface region for conformal plans (both three-dimensional conformal radiotherapy [3DCRT] and intensity-modulated radiotherapy [IMRT]) of 35 breast cancer patients were evaluated and compared with the doses measured with Markus chamber. Materials and Methods: The computed tomography (CT) images of a solid water slab phantom with a Markus chamber (at different depths ranging from 1 mm to 5 mm from the surface) were taken and imported into the TPS. All the conformal treatment plans made in TPS were executed on a linear accelerator and dose agreements between TPS calculated and chamber measured doses were analysed. Results: Results showed that this TPS underestimated the calculated doses in the superficial region by up to 26% and 21%, respectively, with respect to mean and maximum dose values obtained within the effective volume of the chamber used. Conclusion: The uncertainty of doses in the superficial region should be kept in mind when evaluating treatment plans for superficial tumours in TPS.

Indexed as

Breast cancerconformal planning techniquesMarkus chamberskin dosimetrysolid water phantomtreatment planning system

Identifiers

PMID36684697
PMCPMC9847005
OpenAlexW4312655786

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-SA
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.