ArticleThe journal of applied laboratory medicine2023

Comparison of a Novel Thin-Walled 25-gauge Needle Push Button Blood Collection Set to a Standard 23-gauge Needle in a Cancer Patient Population.

Daniel Kirchhoff, Raymond Baser, David Kwong, Lakshmi Ramanathan, Samuel I McCash

Abstract read
In one paragraph

Article in The journal of applied laboratory medicine, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Daniel KirchhoffClinical Chemistry Service, Department of Pathology, Molecular and Cell-Based Medicine, The Mount Sinai Hospital, New York, NY, USA.
Raymond BaserDepartment of Epidemiology & Biostatistics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.ORCID 0000-0002-1644-4090
David KwongClinical Chemistry Service, Department of Pathology and Laboratory Medicine, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Lakshmi RamanathanClinical Chemistry Service, Department of Pathology and Laboratory Medicine, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Samuel I McCashClinical Chemistry Service, Department of Pathology and Laboratory Medicine, Memorial Sloan Kettering Cancer Center, New York, NY, USA.

Funding

X-RAY CRYSTALLOGRAPHYP30CA008748 · NCI · SLOAN-KETTERING INSTITUTE FOR CANCER RES · PI Nicholas D Socci · 1985 to 2026
$347.4M
NCI NIH HHS P30 CA008748
6 · The paper itself

Abstract

backgroundPhlebotomy can be an uncomfortable and even painful experience that increases in likelihood in patients who require frequent blood draws, such as those with cancer. The 25-gauge BD Vacutainer® UltraTouch™ Push Button Blood Collection Set has a smaller outer bore needle size and unique bevel configuration, which in theory should reduce pain associated with phlebotomy. Smaller needles typically cause less pain, however, they have a higher incidence of compromising the specimen integrity. Innovative engineering allows this particular needle to maintain a larger inner bore size similar to a standard 23-gauge needle. This study compares pain and specimen integrity between this novel device and a standard 23-gauge needle.

methodsTwo cohorts of 190 patients each had phlebotomy performed with either needle. Pain experienced was assessed by patient surveys, rating pain from 0 (low) to 10 (high). Sample integrity was assessed by determining the hemolysis index (Abbott Architect c8000).

resultsNo statistically significant difference in median pain scores was observed between the 2 cohorts, P-value: 0.068. The 23-gauge cohort more frequently reported 3+ pain than the 25-gauge cohort, 14/190 vs 5/190. Pain scoring 1 and 2 was more frequent in the 25-gauge cohort, 84/190 vs 54/190. Pain scores of 0 were more frequent in the 23-gauge cohort, 122/190 vs 101/190. This stratification is statistically significant, P-value: 0.003. The 25-gauge needle showed no increase in hemolysis rates, P-value: 0.5.

conclusionsSample integrity was identical between both needles. The 25-gauge needle resulted in less 3+ pain, while pain scoring 1 and 2 was more frequent in the 25-gauge cohort.

Indexed as

HemolysisNeoplasmsHumansNeedlesPainPhlebotomy