Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time.

Jorge Herranz Olazabal, Fokko Wieringa, Evelien Hermeling, Chris Van Hoof

Abstract read
In one paragraph

Article in Bioengineering (Basel, Switzerland), 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Review
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Jorge Herranz OlazabalIMEC, 3000 Leuven, Belgium.ORCID 0000-0002-2347-401X
Fokko WieringaIMEC NL, 5656 AE Eindhoven, The Netherlands.ORCID 0000-0002-7759-7816
Evelien HermelingIMEC NL, 5656 AE Eindhoven, The Netherlands.
Chris Van HoofIMEC, 3000 Leuven, Belgium.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveThe goal was to compare Speckle plethysmography (SPG) and Photoplethysmography (PPG) with non-invasive finger Arterial Pressure (fiAP) regarding Pulse Wave Morphology (PWM) and Pulse Arrival Time (PAT).

methodsHealthy volunteers (n = 8) were connected to a Non-Invasive Blood Pressure (NIBP) monitor providing fiAP pulse wave and PPG from a clinical transmission-mode SpO ANALYSIS: Post-processing was performed in Matlab to obtain SPG and analyze the SPG, PPG and fiAP mean absolute deviations (MADs) on PWM, plus PAT modulation.

resultsAcross 2599 beats, the average fiAP MAD with PPG was 0.17 (0-1) and with SPG 0.09 (0-1). PAT derived from ECG-fiAP correlated as follows: 0.65 for ECG-SPG and 0.67 for ECG-PPG.

conclusionCompared to the clinical NIBP monitor fiAP reference, PWM from an experimental camera-derived non-contact reflective-mode SPG setup resembled fiAP significantly better than PPG from a simultaneously recorded clinical transmission-mode finger clip. For PAT values, no significant difference was found between ECG-SPG and ECG-PPG compared to ECG-fiAP.

Indexed as

camera-basedlaser speckleoptical monitoringPPGspeckle contrast analysisSPG

Identifiers

PMID36671673