Usefulness of large beam-shaping filters at different tube voltages of newborn chest CT.
Article in Physical and engineering sciences in medicine, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Contemporary strategies for dose optimization in pediatric congenital heart disease CT: size-specific acquisition and detector-level advances.Radiological physics and technology · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
12 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundTo investigate optimizing the use of different beam shaping filters (viz. small, medium and large) when using different tube voltages during the newborn chest computed tomography (CT) on a GE Lightspeed VCT scanner.
methodsWe used pediatric anthropomorphic phantoms with a 64 detector-row CT scanner while scanning the chest. A real-time skin dosimeter (RD - 1000; Trek Corporation, Kanagawa, Japan) was positioned into the phantom center of the body, the surface of the body back, and the right and left mammary glands. We performed and compared six scan protocols using small, medium, and large beam shaping filters at 80 and 120 kVp protocols.
resultThere were no significant differences in the image noise for the chest scan among the different beam shaping filters. By using the large beam shaping filter at 80 kVp, it was possible to reduce the exposure dose by 5% in comparison with the small beam shaping filter, and by 10% in comparison with the medium beam shaping filter. By using the large beam shaping filter at 120 kVp, it was possible to reduce the exposure dose by 15% in comparison with the small beam shaping filter and by 20% in comparison with the medium beam shaping filter (p < 0.01).
conclusionThe large beam shaping filter had the most dose reduction effect during newborn chest CT on a GE Lightspeed VCT scanner. The additional copper filtration being present in the large bowtie filter of the GE Lightspeed CT scanner when using different tube voltages is more effective in reducing radiation exposure in children.