Evidence map›Paper›PMID 36621917›Full record

ArticleEuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology2023

A bench study of balloon-expandable valves for the treatment of self-expanding valve failure.

Mariama Akodad, David Meier, Stephanie Sellers, Ole de Backer, Darren Mylotte, Uri Landes, Chris Frawley, Lisa Lynch, Gilbert H L Tang, Lars Sondergaard and 3 more

Open access · greenAbstract read
In one paragraph

Article in EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
5.5field-weighted citation impact, top 3% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed, 25 citations in OpenAlex.

  1. Review
  2. Article
  3. Review
  4. Chimney stenting for preventing coronary obstruction in redo-TAVI with balloon-expandable valves within self-expanding valves.EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2025
    Article
  5. Review
  6. Review
  7. Feasibility of redo-TAVI in the self-expanding ACURATE neo2 valve: a computed tomography study.EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2024
    Article
  8. Impact of leaflet splitting on coronary access after redo-TAVI for degenerated supra-annular self-expanding platforms.EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2024
    Article
  9. Redo-TAVI with the ACURATE neo2 and Prime XL for balloon-expandable transcatheter heart valve failure.EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2024
    Article
  10. Review
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

13 authors at 8 institutions in 4 countries.

Mariama AkodadCentre for Heart Valve Innovation, St Paul's Hospital, University of British Columbia, Vancouver, BC, Canada.
David MeierCentre for Heart Valve Innovation, St Paul's Hospital, University of British Columbia, Vancouver, BC, Canada.
Stephanie SellersCentre for Heart Valve Innovation, St Paul's Hospital, University of British Columbia, Vancouver, BC, Canada.
Ole de BackerDepartment of Cardiology, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.
Darren MylotteDepartment of Cardiology, National University of Ireland, Galway, (NUIG), Galway, Ireland.
Uri LandesEdith Wolfson Medical Center, Holon, Israel and Tel-Aviv University, Tel-Aviv, Israel.
Chris FrawleyBoston Scientific Corporation, Marlborough, MA, USA.
Lisa LynchBoston Scientific Corporation, Marlborough, MA, USA.
Gilbert H L TangDepartment of Cardiovascular Surgery, Mount Sinai Hospital, New York, NY, USA.
Lars SondergaardDepartment of Cardiology, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.
David A WoodCentre for Heart Valve Innovation, St Paul's Hospital, University of British Columbia, Vancouver, BC, Canada.
John G WebbCentre for Heart Valve Innovation, St Paul's Hospital, University of British Columbia, Vancouver, BC, Canada.
Janarthanan SathananthanCentre for Heart Valve Innovation, St Paul's Hospital, University of British Columbia, Vancouver, BC, Canada.
St. Paul's Hospital · CABoston Scientific (United States) · USSt. Paul's Hospital · CACopenhagen University Hospital · DKMount Sinai Hospital · USOllscoil na Gaillimhe – University of Galway · IERigshospitalet · DKUniversity of British Columbia · CA

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundCoronary obstruction and access are concerns in patients undergoing redo transcatheter aortic valve implantation (TAVI).

aimsWe sought to assess the neoskirt height, leaflet overhang, leaflet deflection,and transcatheter heart valve (THV) expansion and performance, at 2 different implant depths, of the SAPIEN 3 Ultra (S3U) within the ACURATE neo2 (ACn2) THV.

methodsAn in vitro study was performed with a 23 mm S3U deployed within a small (S) ACn2 and a 26 mm S3U deployed within a medium (M) and a large (L) ACn2. The S3U outflow was positioned at the top of the ACn2 crown (low implant) and at the base of the commissural post of the ACn2 (high implant). Testing was performed under physiological conditions as per ISO-5840-3 standard.

resultsThe neoskirt height was shorter when the S3U outflow was positioned at a low implantation depth (S: 9.6 mm, M: 12.2 mm, L: 13.8 mm vs S: 15.2 mm, M: 15.1 mm, L: 17.8 mm ACn2 for high implants). Hydrodynamic performance was acceptable for all configurations. Leaflet overhang was <50% for all configurations except the low implant of the 26 mm S3U in the L ACn2 (77.6%). There was a gap from the side of the neoskirt to the outer border of the THV frame which was >2 mm for all configurations. The S3U was underexpanded for all configurations, and the achieved area was 77.9%-92.9% of the expected nominal area.

conclusionsRedo TAVI with an S3U within an ACn2 has favourable hydrodynamics and moderate leaflet overhang. Importantly, the design of the ACn2 results in a neoskirt that is not deflected all the way to the outer dimensions of the THV, hence creating a space that facilitates coronary flow and access.

Indexed as

Aortic Valve StenosisHeart Valve ProsthesisTranscatheter Aortic Valve ReplacementAortic ValveCathetersHumansProsthesis DesignTreatment Outcome

Identifiers

PMID36621917
PMCPMC10174185
OpenAlexW4319777580

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.