Evidence map›Paper›PMID 36532824›Full record

ArticleClinical ophthalmology (Auckland, N.Z.)2022

Post-Market Evaluation of Rotational Stability and Visual Performance of a New Toric Intraocular Lens with Frosted Haptics.

Daniel H Chang, Jerry Hu, Kevin M Miller, Srividhya Vilupuru, Wuchen Zhao

Abstract read
In one paragraph

Article in Clinical ophthalmology (Auckland, N.Z.), 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Daniel H ChangEmpire Eye and Laser Center, Bakersfield, CA, USA.ORCID 0000-0001-6307-6342
Jerry HuTexas Eye & Laser Center, Hurst, TX, USA.
Kevin M MillerDepartment of Ophthalmology, David Geffen School of Medicine at the University of California Los Angeles (UCLA), Los Angeles, CA, USA.ORCID 0000-0002-5128-0685
Srividhya VilupuruJohnson & Johnson Surgical Vision, Inc., Irvine, CA, USA.
Wuchen ZhaoJohnson & Johnson Surgical Vision, Inc., Irvine, CA, USA.ORCID 0000-0001-7662-8177

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: To evaluate 3-month rotational stability, and visual and refractive outcomes of a toric intraocular lens (IOL) with frosted haptics. Patients and Methods: A post-market, prospective, multi-center, single-arm, open-label study conducted at seven clinical sites in the United States. Two hundred and two eyes of 133 subjects with unilateral or bilateral cataracts and corneal astigmatism were implanted with the Tecnis Toric II IOL, Models ZCU150 to 600 (Johnson & Johnson Surgical Vision, Inc.). Lens axis misalignment/rotation, visual acuity, manifest refraction, and surgeon and patient satisfaction were evaluated 3 months postoperatively. Lens rotation was determined with operative and postoperative visit photographs and was analyzed by two independent masked analysts. Results: Mean absolute lens rotation was 0.82° ± 1.00° and 0.94° ± 0.71° at 1 day (n = 189 eyes) and 3 months (n = 185 eyes), respectively. Absolute lens rotation was ≤5° in 98.9% and 100% of eyes at 1 day and 3 months, respectively. At 3 months, postoperative monocular uncorrected and corrected distance visual acuities were 0.004 ± 0.115 LogMAR (20/20) and -0.066 ± 0.092 LogMAR (20/17), mean spherical equivalent was -0.25 D ± 0.35 D and residual refractive cylinder was +0.27 D ± 0.33 D. Surgeons were satisfied/very satisfied with overall clinical outcomes and rotational stability in 99% of eyes, and with uncorrected distance vision in 98.5% of eyes. Conclusion: The study results demonstrate that the Tecnis Toric II IOL has excellent rotational stability with a high percentage of study eyes being within 5° of intended orientation. The study lens demonstrated excellent uncorrected distance visual acuity, reduction of cylinder, and exceptional patient and surgeon satisfaction.

Indexed as

IOL misalignmentlens rotationresidual astigmatismtoric IOL

Identifiers

PMID36532824
PMCPMC9753564

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.