ArticleJournal of cardiovascular electrophysiology2023
In vivo tissue temperatures during 90 W/4 sec-very high power-short-duration (vHPSD) ablation versus ablation index-guided 50 W-HPSD ablation: A porcine study.
Article in Journal of cardiovascular electrophysiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Efficacy and Safety of Pulsed-Field Versus High-Power Short-Duration Ablation for Atrial Fibrillation: A Systematic Review and Meta-Analysis With Reconstructed Time-to-Event Data.Journal of cardiovascular electrophysiology · 2025Pooled it
- Efficacy and Safety of 60 W Versus 90 W, 4-s Radiofrequency Applications to the Esophagus-Adjacent Left Atrial Posterior Wall During Pulmonary Vein Isolation: A Pilot Study.Journal of arrhythmia · 2026Article
- Feasibility and safety of a novel very-high-power short-duration ablation strategy for superior vena cava isolation.Heart rhythm O2 · 2026Article
- Very-high-power, ultra-short-duration radiofrequency for superior vena cava isolation: translational validation in porcine and clinical cohorts.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026Article
- Pulmonary Vein Isolation for Atrial Fibrillation: Comparison of a Very High-Power Very Short-Duration (vHPvSD) Ablation Protocol versus a Hybrid Ablation Approach-Procedural and Mid-Term Outcome Data.Journal of clinical medicine · 2024Article
- Where is the gap after a 90 W/4 s very-high-power short-duration ablation of atrial fibrillation?: Association with the left atrial-pulmonary vein voltage and wall thickness.Journal of arrhythmia · 2024Article
- High-Power Short-Duration Ablation of Paroxysmal and Persistent Atrial Fibrillation.Reviews in cardiovascular medicine · 2023Review
- In vivo tissue temperatures during 90 W/4 sec-very high power-short-duration (vHPSD) ablation versus ablation index-guided 50 W-HPSD ablation: A porcine study.Journal of cardiovascular electrophysiology · 2023Article
Corrections and comments
- Commented on by
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionNeither the actual in vivo tissue temperatures reached with 90 W/4 s-very high-power short-duration (vHPSD) ablation for atrial fibrillation nor the safety and efficacy profile have been fully elucidated.
methodsWe conducted a porcine study (n = 15) in which, after right thoracotomy, we implanted 6-8 thermocouples epicardially in the superior vena cava, right pulmonary vein, and esophagus close to the inferior vena cava. We compared tissue temperatures close to a QDOT MICRO catheter, between during 90 W/4 s-vHPSD ablation during ablation index (AI: target 400)-guided 50 W-HPSD ablation, both targeting a contact force of 8-15 g.
resultsMaximum tissue temperature reached during 90 W/4 s-vHPSD ablation did not differ significantly from that during 50 W-HPSD ablation (49.2 ± 8.4°C vs. 50.0 ± 12.1°C; p = .69) and correlated inversely with distance between the catheter tip and the thermocouple, regardless of the power settings (r = -0.52 and r = -0.37). Lethal temperature (≥50°C) was best predicted at a catheter tip-to-thermocouple distance cut-point of 3.13 and 4.27 mm, respectively. All lesions produced by 90 W/4 s-vHPSD or 50 W-HPSD ablation were transmural. Although there was no difference in the esophageal injury rate (50% vs. 66%, p = .80), the thermal lesion was significantly shallower with 90 W/4 s-vHPSD ablation than with 50W-HPSD ablation (381.3 ± 127.3 vs. 820.0 ± 426.1 μm from the esophageal adventitia; p = .039).
conclusionActual tissue temperatures reached with 90 W/4 s-vHPSD ablation appear similar to those with AI-guided 50 W-HPSD ablation, with the distance between the catheter tip and target tissue being shorter for the former. Although both ablation settings may create transmural lesions in thin atrial tissues, any resulting esophageal thermal lesions appear shallower with 90 W/4 s-vHPSD ablation.
Indexed as