ArticleInternational journal of computer assisted radiology and surgery2023
Steerable catheter based on wire-driven seamless artificial blood vessel tube for endoscopic retrograde transpapillary interventions.
Article in International journal of computer assisted radiology and surgery, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed, 6 citations in OpenAlex.
- Novel steerable-tip catheter successfully used for cannulation of a steeply angled pancreaticojejunostomy anastomotic stenosis.Endoscopy · 2025Article
- Balloon-assisted ERCP for bile duct stones in surgically altered anatomy: current techniques, devices, and evolving strategies.Therapeutic advances in gastroenterology · 2025Review
- Successful reintervention using a novel steerable catheter after dislocation of a transluminal gallbladder stent.Endoscopy · 2024Article
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 1 institution in 1 country.
Funding
Abstract
purposeCurrent steerable catheters (SCs) for endoscopic retrograde cholangiopancreatography (ERCP) have performance limitations caused by an asymmetric multiple-slit tube design with a small maximum bend angle, lesser curvatures, and insufficient durability. We propose a wire-driven SC for balanced bidirectional bending using artificial blood vessel material to overcome these limitations. We assess the SC prototype's steerability using phantom and animal models.
methodsThe SC prototype employed a slit-less and multiple-lumen seamless tube with a polytetrafluoroethylene (PTFE) body with stretch-retractable porous expanded PTFE at the distal end, and loop-formed control wires. We evaluated the wire routing design using a static model. The bending performance was compared with conventional SCs. Feasibility studies were performed, including major duodenal papilla insertions and ductal branch selections in desktop phantoms and a mini-pig model.
resultsThe proposed design reduced the wire contact force by 48% compared to the single wire configuration. The maximum bend angle was 162°, almost twofold larger than that for conventional SCs. The lateral tip position in the bent shape was maximally 56% smaller. The tip flexibility was comparable to conventional SCs, and the insertion resistance was similar to the passive catheters. Phantom studies showed that the SC prototype could perform the large and protuberant papilla insertions and fine ductal branch selections without breaking; the animal study was completed successfully.
conclusionWe propose a wire-driven SC design for ERCP using a multi-lumen seamless tube and two loop-formed control wires, different from the conventional SC design with a multiple-slit tube and single control wire. The SC prototype records balanced bidirectional bending with a maximum bending angle of ± 162° without breakage risk. The phantom and animal studies show that the prototype performance potentially facilitates papilla cannulations and intrahepatic ductal branch seeking.