ArticleIEEE transactions on bio-medical engineering2023
Output Power Computation and Adaptation Strategy of an Electrosurgery Inverter for Reduced Collateral Tissue Damage.
Article in IEEE transactions on bio-medical engineering, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Failure Mechanisms in Off-Label Electrosurgery Devices for BASILICA, LAMPOON, and SESAME.JACC. Cardiovascular interventions · 2026Article
- Electrosurgery: heating, sparking and electrical arcs.Facts, views & vision in ObGyn · 2024Article
- MCU-based Safer Coagulation Mode by Nonfixed Duty Cycle for an Electrosurgery Inverter.IEEE transactions on power electronics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
Abstract
objectiveThis paper investigates two ways of output-power computation, namely, sparse- and multi-sampling-based methods, to overcome sampling speed limitation and arcing nonlinearity for electrosurgery. Moreover, an impedance-based power adaptation strategy is explored for reduced collateral tissue damage.
methodsThe efficacy of the proposed power computation and adaptation strategy are experimentally investigated on a gallium-nitride (GaN)-based high-frequency inverter prototype that allows electrosurgery with a 390 kHz output frequency.
resultsThe sparse-sampling-based method samples output voltage once and current twice per cycle. The achieved power computing errors over 1000 cycles are 1.43 W, 2.54 W, 4.53 W, and 4.89 W when output power varies between 15 W and 45 W. The multi-sampling-based method requires 28 samples of both outputs, and the corresponding errors are 0.02 W, 0.86 W, 1.86 W, and 3.09 W. The collateral tissue damage gauged by average thermal spread is 0.86 mm, 0.43 mm, 1.11 mm, and 0.36 mm for the impedance-based power adaptation against 1.49 mm for conventional electrosurgery.
conclusionBoth power-computation approaches break sampling speed limitations and calculate output power with small errors. However, with arcing nonlinearity presence, the multi-sampling-based method yields better accuracy. The impedance-based power adaptation reduces thermal spreads and diminishes sensor count and cost. SIGNIFICANCE: This paper exemplifies two novel power-computation ways using low-end industrial-scale processors for biomedical research involving high-frequency and nonlinearly distorted outputs. Additionally, this work is the first to present the original impedance-based power adaptation strategy for reduced collateral damage and it may motivate further interdisciplinary research towards collateral-damage-less electrosurgery.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials