ArticleMedical physics2023
High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT.
Article in Medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 17 citations in OpenAlex.
- Cardiac imaging using photon counting CT - benefits, challenges and prospects.The international journal of cardiovascular imaging · 2026Review
- High-pitch photon-counting detector computed tomography angiography of the coronary arteries: Qualitative and quantitative evaluation of monoenergetic image reconstructions.European journal of radiology open · 2025Article
- Performance of iodine quantification through high-pitch dual-source photon-counting CT: a phantom study.Japanese journal of radiology · 2025Article
- Standardizing technical parameters and terms for abdominopelvic photon-counting CT: laying the groundwork for innovation and evidence sharing.Abdominal radiology (New York) · 2024Review
- Cardiac-induced motion of the pancreas and its effect on image quality of ultrahigh-resolution CT.European radiology experimental · 2024Article
- Influence of heart rate and heart rate variability on the feasibility of ultra-fast, high-pitch coronary photon-counting computed tomography angiography.The international journal of cardiovascular imaging · 2023Article
- Photon-Counting Detector Computed Tomography Versus Energy-Integrating Detector Computed Tomography for Coronary Artery Calcium Quantitation.Journal of computer assisted tomographyArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 2 institutions in 1 country.
Funding
Abstract
objectiveTo measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode.
methodsA DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode. Image-based circularity, diameter, and iodine concentration of the iodine cylinder were calculated and compared between the two scanners. With institutional review board approval, in vivo exams were performed with the DS-PCD CT in high-pitch mode. Images were qualitatively compared against patients with similar heart rates that were scanned with DS-EID CT in low-pitch dual-energy mode.
resultsOn iodine maps, the mean circularity was 0.97 ± 0.02 with DS-PCD in high-pitch mode and 0.95 ± 0.06 with DS-EID in low-pitch mode. The mean diameter was 2.9 ± 0.2 mm with DS-PCD and 3.1 ± 0.2 mm with DS-EID, both of which are close to the 3 mm ground truth. For DS-PCD, the mean iodine concentration was 9.6 ± 0.8 mg/ml and this was consistent with the 9.4 ± 0.6 mg/ml value obtained with the cardiac motion disabled. For DS-EID, the concentration was 12.7 ± 1.2 mg/ml with motion enabled and 11.7 ± 0.5 mg/ml disabled. The background noise in the iodine maps was 15.1 HU with DS-PCD and 14.4 HU with DS-EID, whereas the volume CT dose index (CTDI
conclusionIn a cardiac motion phantom, DS-PCD CT can perform accurate material decomposition in high-pitch mode, providing iodine maps with excellent geometric accuracy and robustness to motion at approximately 38% of the dose for similar noise as DS-EID CT.