ArticleJournal of applied clinical medical physics2023

Automatic measurement of beam-positioning accuracy at off-isocenter positions.

Tomohiro Ono, Takahisa Kido, Mitsuhiro Nakamura, Hiraku Iramina, Ryo Kakino, Takashi Mizowaki

Full text read
In one paragraph

Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Tomohiro OnoDepartment of Radiation Oncology and Image-Applied Therapy, Kyoto University, Kyoto, Japan.ORCID https://orcid.org/0000-0001-7398-028X
Takahisa KidoDepartment of Information Technology and Medical Engineering, Human Health Sciences, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
Mitsuhiro NakamuraDepartment of Information Technology and Medical Engineering, Human Health Sciences, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
Hiraku IraminaDepartment of Radiation Oncology and Image-Applied Therapy, Kyoto University, Kyoto, Japan.ORCID https://orcid.org/0000-0003-4059-7932
Ryo KakinoKansai BNCT Medical Center, Osaka, Medical and Pharmaceutical University, Takatsuki, Japan.
Takashi MizowakiDepartment of Radiation Oncology and Image-Applied Therapy, Kyoto University, Kyoto, Japan.ORCID https://orcid.org/0000-0002-8135-8746

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThis study performed an automatic measurement of the off-axis beam-positioning accuracy at a single isocenter via the TrueBeam Developer mode and evaluated the beam-positioning accuracy considering the effect of couch rotational errors.

methodsTrueBeam STx and the Winston-Lutz test-dedicated phantom, with a 3 mm diameter steel ball, were used in this study. The phantom was placed on the treatment couch, and the Winston-Lutz test was performed at the isocenter for four gantry angles (0°, 90°, 180°, and 270°) using an electronic portal imaging device. The phantom offset positions were at distances of 0, 25, 50, 75, and 100 mm from the isocenter along the superior-inferior, anterior-posterior, and left-right directions. Seventeen patterns of multileaf collimator-shaped square fields of 10 × 10 mm

resultsThe mean beam-positioning errors at the isocenter and off-isocenter distances (from the isocenter to ±100 mm) were 0.46-0.60, 0.44-0.91, and 0.42-1.11 mm for the couch angles of 0°, ±0.5°, and ±1°, respectively. The beam-positioning errors increased as the distance from the isocenter and couch rotation increased.

conclusionThese findings suggest that the beam-positioning accuracy at the isocenter and off-isocenter positions can be evaluated quickly and automatically using the TrueBeam Developer mode. The proposed procedure is expected to contribute to an efficient evaluation of the beam-positioning accuracy at off-isocenter positions.

Indexed as

beam-positioning accuracydeveloper modeoff isocenterWinston-Lutz test

Identifiers

PMID36420973