ArticleJournal of applied clinical medical physics2023
A simplified and effective off-axis Winston-Lutz for single-isocenter multi-target SRS.
Article in Journal of applied clinical medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 1 synthesis or guideline pooled it.
- AAPM-RSS Medical Physics Practice Guideline 9.b: SRS-SBRT.Journal of applied clinical medical physics · 2025Guideline
- Distance-adaptive geometric margins for residual rotational uncertainty in single-isocenter multitarget stereotactic radiosurgery.Physics and imaging in radiation oncology · 2026Article
- Clinical implementation of a multifaceted quality assurance phantom for high-precision radiation therapy: an institutional experience.Journal of the Egyptian National Cancer Institute · 2025Article
- Evaluation of artificial intelligence and optical image recognition techniques used in OneIso, an off-axis Winston-Lutz quality assurance phantom.Biomedical physics & engineering express · 2024Article
- Off-iso Winston-Lutz test on seven linear accelerators.Journal of applied clinical medical physics · 2024Article
- Evaluating the Efficacy of Machine Performance Checks as an Alternative to Winston-Lutz Quality Assurance Testing in the TrueBeam Linear Accelerator with HyperArc.Diagnostics (Basel, Switzerland) · 2024Article
- Single isocenter HyperArc treatment of multiple intracranial metastases: Targeting accuracy.Journal of applied clinical medical physics · 2024Article
- A multi-institutional trial evaluating the use of an integrated quality assurance phantom for frameless single-isocenter multitarget stereotactic radiosurgery.Frontiers in oncology · 2024Article
- A simplified and effective off-axis Winston-Lutz for single-isocenter multi-target SRS.Journal of applied clinical medical physics · 2023Article
- Feasibility of Using Machine Performance Check-enhanced Couch Check as a Daily Surrogate for Multisphere Winston-Lutz Testing in Single-isocenter Multitarget Stereotactic Radiosurgery.Journal of medical physicsArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo safely perform single-iso multi-target (SIMT) stereotactic radiosurgery (SRS), clinics must demonstrate SRS delivery accuracy for off-axis targets. The traditional Winston-Lutz (W-L) was widely adopted because it provides a simple and accurate solution for testing radiation-isocenter coincidence that uses a static target, enables testing arbitrary treatment angles, and does not require expensive commercial phantoms. The current noncommercial tests are cumbersome and insufficiently accurate. For an off-axis Winston-Lutz (OAWL) test, one must design MLC fields centered on off-axis targets. Unfortunately, because MLC leaf-interfaces are often misaligned with the target center, accomplishing this presents a nontrivial geometry problem that has not been previously solved in the literature. We present a solution for evaluating SIMT SRS accuracy that provides a straightforward method for creating OAWL test fields and offers all the benefits of the standard W-L test.
methodsWe have developed a method to use any gantry, table, and initial collimator angles to create OAWL fields. This method calculates a series of nested coordinate transformations that produce a small collimator angle adjustment to align the MLC and create a symmetric field around an off-axis target.
resultsFor an 8 cm off-axis target, the described method yields OAWL results within 0.07 mm of standard isocentric W-L results. Our six most recent isocentric W-L tests show max and mean errors of 0.59 and 0.37 mm, respectively. For six runs of our proposed OAWL test, the average max and mean errors are 0.66 and 0.40 mm, respectively.
conclusionThis method accurately evaluates SRS delivery accuracy for off-axis distances that span the majority of a typical human brain for a centered SIMT arc. We have made this method publicly available, so that physicists can employ it within their clinics, foregoing the need for expensive phantoms and improving access to the state-of-the-art SIMT SRS technique.