Evidence map›Paper›PMID 36359572›Full record

ArticleDiagnostics (Basel, Switzerland)2022

T1 Mapping MOLLI 5(3)3 Acquisition Scheme Yields High Accuracy in 1.5 T Cardiac Magnetic Resonance.

Patrick Krumm, Petros Martirosian, Alexander Brendel, Jens M Kübler, Jan M Brendel, Sebastian Gassenmaier, Arne Estler, Meinrad Gawaz, Konstantin Nikolaou, Simon Greulich

Abstract read
In one paragraph

Article in Diagnostics (Basel, Switzerland), 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Effect of the Modified Look-Locker Inversion Recovery Sampling Scheme on Accuracy and Heart Rate Variability in Myocardial T1 Mapping.Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine · 2026
    Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Patrick KrummDepartment of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0003-1705-8439
Petros MartirosianDepartment of Radiology, Section on Experimental Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0001-8383-0883
Alexander BrendelDepartment of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0001-5723-7867
Jens M KüblerDepartment of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0002-3450-7585
Jan M BrendelDepartment of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0002-2729-610X
Sebastian GassenmaierDepartment of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0003-3701-2227
Arne EstlerDepartment of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0002-1930-7497
Meinrad GawazDepartment of Internal Medicine III, Cardiology and Angiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0003-1124-9592
Konstantin NikolaouDepartment of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0003-2668-7325
Simon GreulichDepartment of Internal Medicine III, Cardiology and Angiology, University of Tübingen, 72076 Tübingen, Germany.ORCID 0000-0002-4844-5442

Funding

Deutsche Forschungsgemeinschaft 374031971 - TRR 240
6 · The paper itself

Abstract

Objectives: To systematically compare two modified Look-Locker inversion recovery (MOLLI) T1 mapping sequences and their impact on (1) myocardial T1 values native, (2) post-contrast and (3) extracellular volume (ECV). Methods: 200 patients were prospectively included for 1.5 T CMR for work-up of ischemic or non-ischemic cardiomyopathies. To determine native and post-contrast T1 for ECV calculation, two different T1 mapping MOLLI acquisition schemes, 5(3)3 (designed for native scans with long T1) and 4(1)3(1)2 (designed for post-contrast scans with short T1), were acquired in identical mid-ventricular short-axis slices. Both schemes were acquired in native and post-contrast scans. Results: Datasets from 163 patients were evaluated (age 55 ± 17 years; 38% female). Myocardial T1 native for 5(3)3 was 1017 ± 42 ms vs. 956 ± 40 ms for 4(1)3(1)2, with mean intraindividual difference −61 ms (p < 0.0001). Post-contrast myocardial T1 in patients was similar for both acquisition schemes, with 494 ± 48 ms for 5(3)3 and 490 ± 45 ms for 4(1)3(1)2 and mean intraindividual difference −4 ms. Myocardial ECV for 5(3)3 was 27.6 ± 4% vs. 27 ± 4% for 4(1)3(1)2, with mean difference −0.6 percentage points (p < 0.0001). Conclusions: The T1 MOLLI 5(3)3 acquisition scheme provides a reliable estimation of myocardial T1 for the clinically relevant range of long and short T1 values native and post-contrast. In contrast, the T1 MOLLI 4(1)3(1)2 acquisition scheme may only be used for post-contrast scans according to its designed purpose.