Evidence map›Paper›PMID 36353759›Full record

ArticleEuropace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology2023

Sense-B-noise: an enigmatic cause for inappropriate shocks in subcutaneous implantable cardioverter defibrillators.

Andreas Haeberlin, Haran Burri, Beat Schaer, Pascal Koepfli, Christian Grebmer, Alexander Breitenstein, Tobias Reichlin, Fabian Noti

Registry-linked trialAbstract read
In one paragraph

Article in Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT05713708 (Multicenter Registry for S-ICD Electrodes), which is not on this map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT05713708 unknown statusnot on this map

Multicenter Registry for S-ICD Electrodes

TypeobservationalSponsorUniversitätsklinikum KölnRan2023 to 2024Enrolled1,000ConditionsICD Malfunction, Sudden Cardiac Death
3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. Article
  2. Article
  3. Incidence, implications, and management of sense-B-noise failure in subcutaneous cardioverter-defibrillator patients: insights from a large multicentre registry.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2024
    Article
  4. Article
  5. S-ICD Implantation "Tips and Tricks".Reviews in cardiovascular medicine · 2023
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Andreas HaeberlinDepartment of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.ORCID 0000-0002-9283-0110
Haran BurriUniversity Hospital of Geneva, Cardiology Department, Geneva, Switzerland.ORCID 0000-0002-4393-5338
Beat SchaerDepartment of Cardiology, University Hospital, University of Basel, Basel, Switzerland.ORCID 0000-0003-1768-1968
Pascal KoepfliCantonal Hospital Baden, Baden, Switzerland.
Christian GrebmerDepartment of Cardiology, Kantonsspital Lucerne, Lucerne, Switzerland.ORCID 0000-0002-2927-8087
Alexander BreitensteinDepartment of Cardiology, University Hospital Zurich, Zurich, Switzerland.ORCID 0000-0002-4440-2947
Tobias ReichlinDepartment of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.ORCID 0000-0002-7197-8415
Fabian NotiDepartment of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.ORCID 0000-0002-2719-5149

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

aimsSubcutaneous implantable cardioverter defibrillators (S-ICDs) are well established. However, inappropriate shocks (IAS) remain a source of concern since S-ICDs offer very limited troubleshooting options. In our multicentre case series, we describe several patients who experienced IAS due to a previously unknown S-ICD system issue. METHODS AND

resultsWe observed six patients suffering from this novel IAS entity. The IAS occurred exclusively in primary or alternate S-ICD sensing vector configuration (therefore called 'Sense-B-noise'). IAS were caused by non-physiologic oversensing episodes characterized by intermittent signal saturation, diminished QRS amplitudes, and disappearance of the artefacts after the IAS. Noise/oversensing could not be provoked by manipulation, X-ray did not show evidence for lead/header issues and impedance measurements were within normal limits. The pooled experience of our centres implies that up to ∼5% of S-ICDs may be affected. The underlying root cause was discussed extensively with the manufacturer but remains unknown and is under further investigation.

conclusionSense-B-noise is a novel cause for IAS due to non-physiologic signal oversensing, arising from a previously unknown S-ICD system issue. Sense-B-noise may be suspected if episodes of signal saturation in primary or alternate vector configuration are present, oversensing cannot be provoked, and X-ray and electrical measurements appear normal. The issue can be resolved by reprogramming the device to secondary sensing vector.

Indexed as

Defibrillators, ImplantableHumansDevice failureInappropriate shockOversensingS-ICDSubcutaneous defibrillator

Identifiers

PMID36353759
PMCPMC9935013

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.