Sense-B-noise: an enigmatic cause for inappropriate shocks in subcutaneous implantable cardioverter defibrillators.
Article in Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT05713708 (Multicenter Registry for S-ICD Electrodes), which is not on this map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Multicenter Registry for S-ICD Electrodes
Who cites it
5 citing papers in PubMed.
- The HONEST Cohort Study: Rationale and Design of a Nationwide Subcutaneous Implantable Cardioverter-Defibrillator Cohort.JACC. Advances · 2026Article
- Inappropriate shocks from subcutaneous and transvenous implantable cardioverter-defibrillators: Reports from the FDA MAUDE database.Heart rhythm O2 · 2026Article
- Incidence, implications, and management of sense-B-noise failure in subcutaneous cardioverter-defibrillator patients: insights from a large multicentre registry.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2024Article
- Comparison of Preoperative ECG Screening and Device-Based Vector Analysis in Patients Receiving a Subcutaneous Implantable Cardioverter-Defibrillator.Medicina (Kaunas, Lithuania) · 2023Article
- S-ICD Implantation "Tips and Tricks".Reviews in cardiovascular medicine · 2023Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
aimsSubcutaneous implantable cardioverter defibrillators (S-ICDs) are well established. However, inappropriate shocks (IAS) remain a source of concern since S-ICDs offer very limited troubleshooting options. In our multicentre case series, we describe several patients who experienced IAS due to a previously unknown S-ICD system issue. METHODS AND
resultsWe observed six patients suffering from this novel IAS entity. The IAS occurred exclusively in primary or alternate S-ICD sensing vector configuration (therefore called 'Sense-B-noise'). IAS were caused by non-physiologic oversensing episodes characterized by intermittent signal saturation, diminished QRS amplitudes, and disappearance of the artefacts after the IAS. Noise/oversensing could not be provoked by manipulation, X-ray did not show evidence for lead/header issues and impedance measurements were within normal limits. The pooled experience of our centres implies that up to ∼5% of S-ICDs may be affected. The underlying root cause was discussed extensively with the manufacturer but remains unknown and is under further investigation.
conclusionSense-B-noise is a novel cause for IAS due to non-physiologic signal oversensing, arising from a previously unknown S-ICD system issue. Sense-B-noise may be suspected if episodes of signal saturation in primary or alternate vector configuration are present, oversensing cannot be provoked, and X-ray and electrical measurements appear normal. The issue can be resolved by reprogramming the device to secondary sensing vector.