ArticleZeitschrift fur medizinische Physik2024
Characterizing Off-center MRI with ZTE.
Article in Zeitschrift fur medizinische Physik, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Multi-contrast dual-resolution UTE for myelin fraction mapping and structural imaging.Magma (New York, N.Y.) · 2026Article
- Article
- QuantitativeBone · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo maximize acquisition bandwidth in zero echo time (ZTE) sequences, readout gradients are already switched on during the RF pulse, creating unwanted slice selectivity. The resulting image distortions are amplified especially when the anatomy of interest is not located at the isocenter. We aim to characterize off-center ZTE MRI of extremities such as the shoulder, knee, and hip, adjusting the carrier frequency of the RF pulse excitation for each TR.
methodsIn ZTE MRI, radial encoding schemes are used, where the distorted slice profile due to the finite RF pulse length rotates with the k-space trajectory. To overcome these modulations for objects far away from the magnet isocenter, the frequency of the RF pulse is shifted for each gradient setting so that artifacts do not occur at a given off-center target position. The sharpness of the edges in the images were calculated and the ZTE acquisition with off-center excitation was compared to an acquisition with isocenter excitation both in phantom and in vivo off-center MRI of the shoulder, knee, and hip at 1.5 and 3T MRI systems.
resultsDistortion and blurriness artifacts on the off-center MRI images of the phantom, in vivo shoulder, knee, and hip images were mitigated with off-center excitation without time or noise penalty, at no additional computational cost.
conclusionThe off-center excitation allows ZTE MRI of the shoulder, knee, and hip for high-bandwidth image acquisitions for clinical settings, where positioning at the isocenter is not possible.