Performance evaluation of Hipee S2 point-of-care testing urine dipstick analyser: a cross-sectional study.
Article in BMJ open, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 3 citations in OpenAlex.
- MyACR: A Point-of-Care Medical Device for Determination of Albumin-Creatinine Ratio (uACR) in Random Urine Samples as a Marker of Nephropathy.Diagnostics (Basel, Switzerland) · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 4 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveWith advances in mobile technology, smartphone-based point-of-care testing (POCT) urinalysis hold great potential for disease screening and health management for clinicians and individual users. The purpose of this study is to evaluate the analytical performance of Hipee S2 POCT urine dipstick analyser.
designA multicentre, hospital-based, cross-sectional study.
settingAnalytical performance of the POCT analyser was conducted at a clinical laboratory, and method comparison was performed at three clinical laboratories in China.
participantsUrine samples were collected from 1603 outpatients and inpatients at three hospitals, and 5 health check-up population at one of the hospitals. OUTCOME MEASURES: All tests were performed by clinical laboratory technicians. Precision, drift, carry-over, interference and method comparison of Hipee S2 were evaluated. Diagnostic accuracy of semiquantitative albumin-to-creatinine ratio (ACR) for albuminuria was carried out using quantitative ACR as the standard.
resultsThe precision for each parameter, assessed by control materials, was acceptable. No sample carry-over or drift was observed. Ascorbate solution with 1 g/L had an inhibitory effect for the haemoglobin test. Agreement for specific gravity (SG) varied between moderate to substantial (κ values 0.496-0.687), for pH was moderate (κ values 0.423-0.569) and for other parameters varied between substantial to excellent (κ values 0.669-0.991), on comparing the Hipee S2 with laboratory analysers. The semiquantitative microalbumin and creatinine were highly correlated with the quantitative results. The sensitivity of semiquantitative ACR to detect albuminuria was 87.2%-90.7%, specificity was 70.7%-78.4%, negative predictive value was 85.3%-87.9% and positive predictive value was 73.9%-83%.
conclusionsHipee S2 POCT urine analyser showed acceptable analytical performance as a semiquantitative method. It serves as a convenient alternate device for clinicians and individual users for urinalysis and health management. In addition, the POCT semiquantitative ACR would be useful in screening for albuminuria.